The Utility of Somatosensory Evoked Potential Monitoring During Cervical Spine Surgery: How Often Does It Prompt Intervention and Affect Outcome?
نویسندگان
چکیده
STUDY DESIGN Retrospective review of the results of somatosensory evoked potentials (SSEP) performed in cervical spine surgery. PURPOSE To evaluate the utility of spinal cord monitoring during cervical spine surgery in a single surgeon's practice, based on how often it prompted an intraoperative intervention. OVERVIEW OF LITERATURE Intraoperative monitoring during cervical spine surgery is not a universally accepted standard of care. This is due in part to the paucity of literature regarding the impact of monitoring on patient management or outcome. METHODS SSEP for tibial, median, and ulnar nerves were monitored in 809 consecutive cervical spine operations performed by a single surgeon. The average patient age was 52 years (range, 2 to 88 years), with 472 males and 339 females. Cases were screened for significant degradation or loss of SSEP data. Specific attention was paid to 1) what interventions were performed in response to the SSEP degradation with subsequent improvement, and 2) whether SSEP changes corresponded with postoperative neurological deficits. RESULTS Seventeen of 809 patients (2.1%) had SSEP degradation that met warning criteria and therefore prompted intervention. Release of shoulder tape (8) or traction (4) most often resulted in SSEP improvement. Failure of SSEP data to return to within acceptable limits of baseline was associated with neurological deficit (p=0.04). Two patients awoke with new postoperative neurological deficits, which resolved in 6 hours and 2 months respectively. Patients with ossification of the posterior longitudinal ligament (OPLL) were at seven-fold greater risk of intraoperative SSEP degradation. CONCLUSIONS SSEP monitoring in this surgical population proved sensitive to perioperative factors which may increase the risk of postoperative neurologic deficit, and probably prevented neurological deficits in 15 of 809 patients (1.9%). Improvement in data following intervention appears to correlate well with unchanged neurologic status. Experience with intraoperative monitoring in this patient series has led to incorporation of these techniques as a standard of care in cervical spine surgeries performed by this surgeon.
منابع مشابه
Comparison of transcranial electric motor and somatosensory evoked potential monitoring during cervical spine surgery.
BACKGROUND There has been little enthusiasm for somatosensory evoked potential monitoring in cervical spine surgery as a result, in part, of the increased risk of motor tract injury at this level, to which somatosensory monitoring may be insensitive. Transcranial electric motor evoked potential monitoring allows assessment of the motor tracts; therefore, we compared transcranial electric motor ...
متن کاملIntraoperative somatosensory evoked potential monitoring during anterior cervical discectomy and fusion in nonmyelopathic patients--a review of 1,039 cases.
BACKGROUND CONTEXT Intraoperative somatosensory evoked potential (SSEP) monitoring has been shown to reduce the incidence of new postoperative neurological deficits in scoliosis surgery. However, its usefulness during cervical spine surgery remains a subject of debate. PURPOSE To determine the utility of intraoperative SSEP monitoring in a specific patient population (those with cervical radi...
متن کاملNeurophysiological monitoring of spinal cord function during instrumented anterior cervical fusion.
BACKGROUND CONTEXT Somatosensory evoked potentials (SSEPs) monitor global spinal cord function, and the interpretation of motor loss is based on inferred rather than direct measurements. Therefore, SSEPs may not be useful for identifying motor function deficits caused by anterior spinal column injury or nerve root injury during decompression or placement of instrumentation. For these reasons, a...
متن کاملEvolving compartment syndrome detected by loss of somatosensory- and motor-evoked potential signals during cervical spine surgery.
Neurologic injury is a rare but devastating complication of spinal surgery that can result in mild sensory to severe motor deficits. Surgeons increasingly use electrophysiological spinal cord function monitoring, including somatosensory- and motor-evoked potentials, intraoperatively to provide information about spinal cord function, aid in surgical decision making, improve outcomes, and reduce ...
متن کاملBilateral C5 Palsy Following a Circumferential Surgery for Cervical Spondylotic Myelopathy: Case Report and Review (In press)
Background and Importance: The development of C5 root palsy is a well-known potential complication of cervical spine surgeries for the correction of cervical spondylotic myelopathy. It typically occurs unilaterally but on extremely rare occasions it might occur as a bilateral anomaly. The value of intraoperative neurophysiologic monitoring in detection of iatrogenic acute versus delayed onset C...
متن کامل