Reference photon dosimetry data: A preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator
نویسندگان
چکیده
The dosimetric characteristics for modern computer-controlled linear accelerators with the same make, model, and nominal energy are known to be very similar, as long as the machines are unaltered from the manufacturer's original specifications. In this preliminary study, a quantitative investigation of the similarity in the basic photon dosimetry data from the Siemens Primus linear accelerators at eight different institutions is reported. The output factor, percentage depth dose (PDD), and in-air off-axis factor (OAF) for the 6 and 18 MV photon beams measured or verified by the Radiological Physics Center (RPC) were analyzed. The RPC-measured output factors varied by less than about 2% for each field size. The difference between the maximum and minimum RPC-verified PDD values at each depth was less than about 3%. The difference between the maximum and minimum RPC-measured in-air OAF was no more than 4% at all off-axis distances considered in this study. These results strongly suggest that it is feasible to establish a reference photon dosimetry data set for each make, model, and nominal energy, universally applicable to those machines unaltered from the manufacturers' original specifications, within a clinically acceptable tolerance (e.g., approximately +/-2%).
منابع مشابه
Monte Carlo Simulation of Siemens Primus plus Linac for 6 and 18 MV Photon Beams
Objective: The aim of the present study is to simulate 6 MV and 18 MV photon beam energies of a Siemens Primus Plus medical linear accelerator (Linac) and to verify the simulation by comparing the results with the measured data.Methods: The main components of the head of Siemens Primus Plus linac were simulated using MCNPX Monte Carlo (MC) code. To verify the results, experimental data of perce...
متن کاملCharacterization of Wedge Factors and Dose Distributions in Radiotherapy with Symmetric and Asymmetric Physical Wedged Beams of 6 MV Photon Beam
Introduction: Physical wedge by modify photon beam shape and intensity has been utilized in radiotherapy to obtain uniformly dose distribution in tumor site with reduced hot spots. Calculation of dosimetric parameters for both symmetric and asymmetric wedged fields is proved necessary during linear accelerator (Linac) commissioning. The present study aimed to achieve o...
متن کاملQuantification the dosimetric parameters of asymmetric physical wedged-6MV photon beam
Introduction: Physical wedge as a useful tool has been utilized in radiotherapy to modify photon beam shape and intensity such that it distributes dose uniformly in tumor site and reduces hot points. Since during Linac commissioning dosimetric parameters like output factors and lateral dose profiles are measured only for symmetric open and wedged fields, so calculation the par...
متن کاملMeasurement Of Scatter Factors For Small Photon Fields Using Gaf chromic EBT2 Film
Introduction: Small field dosimetry is challenging for radiotherapy dosimetry. measurement of the output factor in the air and water (Sc, Scp) is one of the input parameters for commissioning of treatment planning systems and beam modeling. The aims of this study are to measured Sc,Scp for small fields with EBT2 and Ion chamber and design a appropriate mini-phantom for small fi...
متن کاملSiemens primus accelerator simulation using EGSnrc Monte Carlo code and gel dosimetry validation with optical computed tomography system by EGSnrc code
Monte Carlo method is the most accurate method for simulation of radiation therapy equipment. The linear accelerators (linac) are currently the most widely used machines in radiation therapy centers. Monte Carlo modeling of the Siemens Primus linear accelerator in 6 MeV beams was used. Square field size of 10 × 10 cm2 produced by the jaws was compared with TLD. Head simulation of Siemens accele...
متن کامل