A simulator study of adverse wear with metal and cement debris contamination in metal-on-metal hip bearings

نویسندگان

  • T. Halim
  • I. C. Clarke
  • M. D. Burgett-Moreno
  • T. K. Donaldson
  • C. Savisaar
  • J. G. Bowsher
چکیده

OBJECTIVES Third-body wear is believed to be one trigger for adverse results with metal-on-metal (MOM) bearings. Impingement and subluxation may release metal particles from MOM replacements. We therefore challenged MOM bearings with relevant debris types of cobalt-chrome alloy (CoCr), titanium alloy (Ti6Al4V) and polymethylmethacrylate bone cement (PMMA). METHODS Cement flakes (PMMA), CoCr and Ti6Al4V particles (size range 5 µm to 400 µm) were run in a MOM wear simulation. Debris allotments (5 mg) were inserted at ten intervals during the five million cycle (5 Mc) test. RESULTS In a clean test phase (0 Mc to 0.8 Mc), lubricants retained their yellow colour. Addition of metal particles at 0.8 Mc turned lubricants black within the first hour of the test and remained so for the duration, while PMMA particles did not change the colour of the lubricant. Rates of wear with PMMA, CoCr and Ti6Al4V debris averaged 0.3 mm(3)/Mc, 4.1 mm(3)/Mc and 6.4 mm(3)/Mc, respectively. CONCLUSIONS Metal particles turned simulator lubricants black with rates of wear of MOM bearings an order of magnitude higher than with control PMMA particles. This appeared to model the findings of black, periarticular joint tissues and high CoCr wear in failed MOM replacements. The amount of wear debris produced during a 500 000-cycle interval of gait was 30 to 50 times greater than the weight of triggering particle allotment, indicating that MOM bearings were extremely sensitive to third-body wear. Cite this article: Bone Joint Res 2015;4:29-37.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The increase in cobalt release in metal-on-polyethylene hip bearings in tests with third body abrasives

Hypersensitivity reactions in patients receiving metal-on-metal hip replacements have been attributed to corrosion products as observed by elevated cobalt and chromium ions in the blood. Although the majority of cases are reported in metal-on-metal, incidences of these reactions have been reported in the metal-on-polyethylene patient population. To date, no in vitro study has considered cobalt ...

متن کامل

Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear.

Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clin...

متن کامل

Influence of simulator kinematics on the wear of metal-on-metal hip prostheses.

There is now considerable interest in metal-on-metal bearings for hip prostheses. Extremely low wear rates (0.1 mm3/10(6) cycles) have been reported in some simulator studies, while in vivo studies, although still very low, have shown wear rates of the order of 1 mm3/10(6) cycles. The aim of this study was to compare wear rates of metal-on-metal bearings in two hip simulators with different kin...

متن کامل

The effect of metal ions in solution on bacterial growth compared with wear particles from hip replacements.

The biological significance of cobalt-chromium wear particles from metal-on-metal hip replacements may be different to the effects of the constituent metal ions in solution. Bacteria may be able to discriminate between particulate and ionic forms of these metals because of a transmembrane nickel/cobalt-permease. It is not known whether wear particles are bacteriocidal. We compared the doubling ...

متن کامل

CoCrMo metal-on-metal hip replacements.

After the rapid growth in the use of CoCrMo metal-on-metal hip replacements since the second generation was introduced circa 1990, metal-on-metal hip replacements have experienced a sharp decline in the last two years due to biocompatibility issues related to wear and corrosion products. Despite some excellent clinical results, the release of wear and corrosion debris and the adverse response o...