Clinical assessment and characterization of a dual‐tube kilovoltage X‐ray localization system in the radiotherapy treatment room
نویسندگان
چکیده
INTRODUCTION Although flat-panel based X-ray imaging has been well tested in diagnostic radiology, its use as an image-guided-radiotherapy (IGRT) system in a treatment room is new and requires systematic assessment. MATERIALS AND METHODS BrainLab Novalis IGRT system was used for this study. It consists of two floor mounted kV X-ray tubes projecting obliquely into two flat-panel detectors mounted on the ceiling. The system automatically fuses the 2D localization images with 3D simulation CT image to provide positioning guidance. The following characteristics of the system were studied: (1) Coincidence of the isocenters between the IGRT and Linac; (2) Image quality; (3) Exposure; (4) Linearity, uniformity and repeatability. RESULTS (1) Localization accuracy and coincidence of the isocenters between the IGRT and Linac was better than 1-mm. (2) The spatial resolution was quantified using the relative modulation-transfer-function with f50=0.7-0.9 lp/mm. The variation of contrast-noise-ratio with technical settings was measured. (3) The maximal exposure of an image was less than 95 mR. An empirical relation between the exposure and the X-ray technical setting was derived. (4) The linearity, uniformity and repeatability of the system generally meet the requirements. CONCLUSION The system can be safely and reliably used as a target localization device.
منابع مشابه
Chest-wall irradiation by kilovoltage x-rays: Observed local control rate versus predicted radiobiological tumor control probability
Introduction: The high local control rates observed in patients with breast cancer who had been treated by modified radical mastectomy, chemotherapy and chest-wall irradiation using 120 kV X-rays (50 Gy in 25 fractions) motivated us to investigate their observed tumor control probability (TCP) and test a well-established radiobiological TCP model against our data. Mat...
متن کاملAdvances in kilovoltage x-ray beam dosimetry.
This topical review provides an up-to-date overview of the theoretical and practical aspects of therapeutic kilovoltage x-ray beam dosimetry. Kilovoltage x-ray beams have the property that the maximum dose occurs very close to the surface and thus, they are predominantly used in the treatment of skin cancers but also have applications for the treatment of other cancers. In addition, kilovoltage...
متن کاملQuality Control Assessment of Conventional Radiology Devices in Iran
Introduction Quality control (QC) techniques are used in monitoring and maintenance of the components of an x-ray system. QC of radiology devices plays a significant role in reduction of medication dose and optimization of image quality. This study aimed to conduct QC tests on randomly selected radiology devices, installed in diagnostic imaging departments of Iran. Materials and Methods In tota...
متن کاملRadiosensitization effect of ZnO nanoparticles in lung cancer cells at clinically relevant megavoltage energy
Introduction: Radiation therapy is one of the major modalities that have long been used in cancer treatment. Radiotherapy is often accompanied by early and late toxicity and side effects and narrow therapeutic window. Similarity in radiation absorption properties of tumors and neighboring healthy tissues is often the reason for low specificity of radiation therapy. Development ...
متن کاملQuality assessment of conventional X-ray diagnostic equipment by measuring X-ray exposure and tube output parameters in Great Khorasan Province, Iran
Introduction: Regular implementation of quality control (QC) program in diagnostic X-ray facilities may affect both image quality and patient radiation dose due to the changes in exposure parameters. Therefore, this study aimed to investigate the status of randomly selected conventional radiographic X-ray devices installed in radiology centers of Great Khorasan Province, Iran, to produce the da...
متن کامل