OBJECTIVE Modified anterior tension wiring with K-wires and cannulated lag screws with anterior tension wiring are currently the fixation of choice for patellar fractures. Failure of fixation, migration of the wires, postoperative pain and resulting revision surgery, however, are not uncommon. After preliminary biomechanical testing of a new fixed-angle plate system especially designed for fixation of patella fractures the aim of this study was to evaluate the surgical and anatomical feasibility of implanting such a plate-device at the human patella. METHODS In six fresh unfixed female cadavers without history of previous fractures around the knee (average age 88.8 years) a bilateral fixed-angle plate fixation of the patella was carried out after previous placement of a transverse central osteotomy. Operative time, intraoperative problems, degree of retropatellar arthritis (following Outerbridge), quality of reduction and existence of any intraarticular screw placement have been raised. In addition, lateral and anteroposterior radiographs of all specimens were made. RESULTS Due to the high average age of 88.8 years no patella showed an unimpaired retropatellar articular surface and all were severely osteoporotic, which made a secure fixation of the reduction forceps during surgery difficult. The operation time averaged 49 minutes (range: 36-65). Although in postoperative X-rays the fracture gap between the fragments was still visible, the analysis of the retropatellar surface showed no residual articular step or dehiscence > 0.5 mm. Also in a total of 24 inserted screws not one intraarticular malposition was found. No intraoperative complications were noticed. CONCLUSIONS Osteosynthesis of a medial third patella fracture with a bilateral fixed-angle plate-device is surgically and anatomically feasible without difficulties. Further studies have to depict whether the bilateral fixed-angle plate-osteosynthesis of the patella displays advantages over the established operative procedures.
منابع مشابه
Fixation of multifragmentary patella fractures using a bilateral fixed-angle plate.
This biomechanical study is the first to compare 3 fixation methods-bilateral fixed-angle plate, modified anterior tension wiring, and cannulated lag screws with anterior tension wiring-in multifragmentary distal patella fractures. A T-shaped 3-part fracture simulating a multifragmentary articular distal patella fracture (AO/OTA 34-C2.2) was created in 18 human cadaver knee specimens. Three gro...
متن کاملTreatment of posterolateral tibial plateau fractures with modified Frosch approach: a cadaveric study and case series
This study aimed to investigate the surgical techniques and the clinical efficacy of a modified Frosch approach in the treatment of posterolateral tibial plateau fractures. The standard Frosch approach was performed on 5 fresh-frozen cadavers. The mean bony surface area was measured upon adequate exposure of the proximal tibial cortex. Lateral proximal tibial plate and posterolateral T-plates w...
متن کاملA Pitfall in Fixation of Distal Humeral Fractures with Pre-Contoured Locking Compression Plate
Anatomically precontoured locking plates are intended to facilitate the fixation of articular fractures and particularly those associated with osteoporosis. Fractures of the distal humerus are relatively uncommon injuries where operative intervention can be exceptionally challenging. The distal humeral trochlea provides a very narrow anatomical window through which to pass a fixed-angle locking...
متن کاملLocked plate stabilization of problem fractures.
http://dx.doi.org/10.3415/VCOT-15-12-0202 Locking plates of various designs and specification are readily available commercially for fracture stabilization. All of these implants perform similarly, acting as angle stable internal fixators in regard to the plate-screw junction. There is a range of different mechanisms employed to achieve an angle stable locking connection between the plate and t...
متن کاملThe Use of Dual Reconstruction Plates for Failed Fixation of Patellar Fracture after Total Knee Replacement: A Case Report
Patellar fracture after total knee replacement (TKR) is one of the challenging problems in periprosthetic fracture. Open reduction with internal fixation (ORIF), as tension band wiring (TBW), usually required in cases with extensor mechanism disruption. However, many studies reported a high failure rate after using this technique. In this report, we presented an interesting case of periprosthet...