The effect of acrylic nails on the measurement of oxygen saturation as determined by pulse oximetry.

نویسنده

  • S M Peters
چکیده

Pulse oximetry (Spo2) is a simple, noninvasive method that is widely used to determine oxygen saturation in patients undergoing surgical procedures. Artificial acrylic nails have recently become fashionable to strengthen and lengthen nails. This study investigates the effect of unpolished acrylic nails on the measurement of oxygen saturation by pulse oximetry. Data were collected during a 3-month period. Thirty women, average age 32 years (range 18 to 61 years), were recruited at a high-volume nail salon in northwestern Pennsylvania. A baseline pulse oximetry reading was obtained on each subject's natural, unpolished fingernail using a Nellcor N-20/N-20P portable pulse oximeter (Nellcor Incorporated, Hayward, California). A licensed nail technician applied the acrylic compound to the same finger. After the compound had hardened in approximately 5 minutes, a second reading was obtained on the unpolished acrylic nail. The mean pulse oximetry reading at baseline was 97.33% and after acrylic nail application, 97.58%. Using a paired Student's t test, no statistically significant differences existed between readings. This study demonstrates that unpolished acrylic nails do not affect pulse oximetry measurements of oxygen saturation. Patients may not need to remove unpolished acrylic nails before surgery.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of Nail Polish on Pulse Oximetry Findings in Healthy Volunteers: A Randomized Clinical Trail

Background and Objectives: Pulse oximetry is the most common technique for monitoring hemoglobin oxygen saturation (SpO2). Different colors and brands of nail polish may cause disturbance in the reading and interpretation of oxygen saturation. The aim of this study was to determine the effect of different colors of nail polish on oxygen saturation measured by pulse oximeter. Methods: Thirty he...

متن کامل

Comparative study of pulse oximetry, physical examination and echocardiography results in the diagnosis of congenital heart defects in neonates in the first 24 hours of life

Introduction: Congenital heart disease is the most common congenital anomaly. About 50% of Neonates with congenital heart disease are asymptomatic in the first few days of life and are not diagnosed on initial examination. Pulse oximetry is a non-invasive method that can show the percentage of oxygen saturation in the blood and congenital heart disease. Methods: This was a descriptive cross-se...

متن کامل

بررسی نتایج پالس اکسی‌متری دست و پا در تشخیص زودهنگام بیماری‌های قلبی مادرزادی سیانوتیک در نوزادان متولد شده در بیمارستان شبیه خوانی کاشان در 6 ماهه اول سال 1385

  Background : Delayed diagnosis of critical and cyanotic congenital heart diseases in asymptomatic newborns can lead to significant morbidity and mortality in this age group. The aim of this study was to determine the utility of hand and foot oxygen saturation as a screening test for the early detection of critical and cyanotic congenital heart diseases in asymptomatic newborns in the first da...

متن کامل

Evaluation of Pulse Oximetry in the Early Detection of Congenital Heart Diseases in Newborns

Background: Congenital heart diseases, which are asymptomatic at birth, are the most important causes of infant mortality. This study aimed to evaluate the role of pulse oximetry in the early detection of congenital heart diseases among newborns.Methods: In this cross-sectional descriptive study, 1230 newborns who were born in university hospitals in an urban area of Iran were placed unde...

متن کامل

Relationship Between Arterial Oxygen Saturation and Hematocrit, and Effect of Slow Deep Breathing on Oxygen Saturation in Himalayan High Altitude Populations

lIn the cross-sectional study of two populations residing at altitude of 2800 m and 3760 m are compared for the difference in hematocrit. The oxygen carrying capacity of arterial haemoglobin (SaO2) is determined by pulse oximetry. The sample is drawn from the natives of two small villages, Thini at Jomsom (2800 m) and Jharkot (3760 m) in Mustang district of Nepal. The natives at 2800 m are term...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • AANA journal

دوره 65 4  شماره 

صفحات  -

تاریخ انتشار 1997