Effects of voxel size and iterative reconstruction parameters on the spatial resolution of 99mTc SPECT/CT
نویسنده
چکیده
The purpose of this study was to evaluate the effects of voxel size and iterative reconstruction parameters on the radial and tangential resolution for 99mTc SPECT as a function of radial distance from isocenter. SPECT/CT scans of eight coplanar point sources of size smaller than 1 mm3 containing high concentration 99mTc solution were acquired on a SPECT/CT system with 5/8 inch NaI(Tl) detector and low-energy, high-resolution collimator. The tomographic projection images were acquired in step-and-shoot mode for 360 views over 360° with 250,000 counts per view, a zoom of 2.67, and an image matrix of 256 × 256 pixels that resulted in a 0.9 × 0.9 × 0.9 mm3 SPECT voxel size over 230 mm field-of-view. The projection images were also rebinned to image matrices of 128 × 128 and 64 × 64 to yield SPECT voxel sizes of 1.8 × 1.8 × 1.8 and 3.6 × 3.6 × 3.6 mm3, respectively. The SPECT/CT datasets were reconstructed using the vendor-supplied iterative reconstruction software that incorporated collimator-specific resolution recovery, CT-based attenuation correction, and dual-energy window-based scatter correction using different combinations of iterations and subsets. SPECT spatial resolution was estimated as the full width at half maximum of the radial and tangential profiles through the center of each point source in reconstructed SPECT images. Both radial and tangential resolution improved with higher iterations and subsets, and with smaller voxel sizes. Both radial and tangential resolution also improved with radial distance further away from isocenter. The magnitude of variation decreased for smaller voxel sizes and for higher number of iterations and subsets. Tangential resolution was found not to be equal to the radial resolution, and the nature of the anisotropy depended on the distribution of the radionuclide and on the reconstruction parameters used. The tangential resolution converged faster than the radial resolution, with higher iterations and subsets. SPECT resolution was isotropic and independent of radial distance when reconstructed using filtered back-projection. SPECT spatial resolution and therefore quantification of SPECT uptake via partial-volume correction in clinical images were found to depend on the nature of activity distribution within the SPECT field-of-view and on the specific choice of iterative reconstruction parameters.
منابع مشابه
Impact of Novel Incorporation of CT-based Segment Mapping into a Conjugated Gradient Algorithm on Bone SPECT Imaging: Fundamental Characteristics of a Context-specific Reconstruction Method
Objective(s): The latest single-photon emission computed tomography (SPECT)/computed tomography (CT) reconstruction system, referred to as xSPECT Bone™, is a context-specific reconstruction system utilizing tissue segmentation information from CT data, which is called a zone map. The aim of this study was to evaluate theeffects of zone-map enhancement incorporated into the ordered-subset conjug...
متن کاملOptimization of an ultra-high-resolution rectangular pixelated parallel-hole collimator with a CZT pixelated semiconductor detector for HiRe-SPECT system
Introduction: In nuclear medicine, the use of a pixelated semiconductor detector such as CZT is an of growing interest for introducing new devices. Especially, the spatial resolution can be improved by using a pixelated parallel-hole collimator with equal holes and pixel sizes based on the pixelated detector. The purpose of this study was to compare the effect of pixelated and ...
متن کاملEvaluation of iterative reconstruction method and attenuation correction on brain dopamine transporter SPECT using anthropomorphic striatal phantom
Objective(s): The aim of this study was to determine the optimal reconstruction parameters for iterative reconstruction in different devices and collimators for dopamine transporter (DaT) single-photon emission computed tomography (SPECT). The results were compared between filtered back projection (FBP) and different attenuation correction (AC) methods.Methods: An anthropomorphic striatal phant...
متن کاملImpact of reconstruction method on quantitative parameters of 99mTc-TRODAT-1 SPECT
Introduction: Quantitative evaluation is recommended to improve diagnostic ability and serial assessment of dopamine transporter (DAT) density scans. We decided to compare the ordered subsets expectation-maximization (OSEM) with filtered back-projection (FBP), and to investigate the impact of different iteration and cut-off frequencies on SBR values. Methods</stro...
متن کاملThe effect of focal spot size on the spatial resolution of variable resolution X-ray CT scanner
Background: A variable resolution X-ray (VRX) CT scanner provides a great increase in the spatial resolution. In VRX CT scanners, the spatial resolution of the system and its field of view (FOV) can be changed according to the object size. One of the main factors that limit the spatial resolution of VRX CT scanner is the effect of the X-ray focal spot. Materials and Methods: A theoreti...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 12 شماره
صفحات -
تاریخ انتشار 2011