Are component positioning and prosthesis size associated with hip resurfacing failure?
نویسندگان
چکیده
BACKGROUND Recent studies suggest that there is a learning curve for metal-on-metal hip resurfacing. The purpose of this study was to assess whether implant positioning changed with surgeon experience and whether positioning and component sizing were associated with implant longevity. METHODS We evaluated the first 361 consecutive hip resurfacings performed by a single surgeon, which had a mean follow-up of 59 months (range, 28 to 87 months). Pre and post-operative radiographs were assessed to determine the inclination of the acetabular component, as well as the sagittal and coronal femoral stem-neck angles. Changes in the precision of component placement were determined by assessing changes in the standard deviation of each measurement using variance ratio and linear regression analysis. Additionally, the cup and stem-shaft angles as well as component sizes were compared between the 31 hips that failed over the follow-up period and the surviving components to assess for any differences that might have been associated with an increased risk for failure. RESULTS Surgeon experience was correlated with improved precision of the antero-posterior and lateral positioning of the femoral component. However, femoral and acetabular radiographic implant positioning angles were not different between the surviving hips and failures. The failures had smaller mean femoral component diameters as compared to the non-failure group (44 versus 47 millimeters). CONCLUSIONS These results suggest that there may be differences in implant positioning in early versus late learning curve procedures, but that in the absence of recognized risk factors such as intra-operative notching of the femoral neck and cup inclination in excess of 50 degrees, component positioning does not appear to be associated with failure. Nevertheless, surgeons should exercise caution in operating patients with small femoral necks, especially when they are early in the learning curve.
منابع مشابه
Performance of the resurfaced hip. Part 1: the influence of the prosthesis size and positioning on the remodelling and fracture of the femoral neck.
Hip resurfacing is an established treatment for osteoarthritis in young active patients. Failure modes include femoral neck fracture and prosthesis loosening, which may be associated with medium-term bone adaptation, including femoral neck narrowing and densification around the prosthesis stem. Finite element modelling was used to indicate the effects of prosthesis sizing and positioning on the...
متن کاملThe effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip.
Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive pati...
متن کاملHip Resurfacing Arthroplasty: Investigating the Femoral Component in the Sagittal Plane
The outcomes of hip resurfacing arthroplasty are largely dependent on prosthesis positioning. The biomechanics of notching, accurate measurement of femoral implant version and the use of computer navigation of the Birmingham Hip Resurfacing procedure were studied in this work. First, biomechanical tests were conducted with varying notch sizes and femoral positions, and it was determined that an...
متن کاملEarly failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear.
Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clin...
متن کاملSex differences in the morphological failure patterns following hip resurfacing arthroplasty
BACKGROUND Metal-on-metal hybrid hip resurfacing arthroplasty (with a cementless acetabular component and a cemented femoral component) is offered as an alternative to traditional total hip arthroplasty for the young and active adult with advanced osteoarthritis. Although it has been suggested that women are less appropriate candidates for metal-on-metal arthroplasty, the mechanisms of prosthes...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 11 شماره
صفحات -