Keratoconus diagnosis using Corvis ST measured biomechanical parameters
نویسندگان
چکیده
PURPOSE To assess the diagnostic power of the Corneal Visualization Scheimpflug Technology (Corvis ST) provided corneal biomechanical parameters in keratoconic corneas. METHODS The following biomechanical parameters of 48 keratoconic eyes were compared with the corresponding ones in 50 normal eyes: time of the first applanation and time from start to the second applanation [applanation-1 time (A1T) and applanation-2 time (A2T)], time of the highest corneal displacement [highest concavity time (HCT)], magnitude of the displacement [highest concavity deformation amplitude (HCDA)], the length of the flattened segment in the applanations [first applanation length (A1L) and second applanation length (A2L)], velocity of corneal movement during applanations [applanation-1 velocity (A1V) and applanation-2 velocity (A2V)], distance between bending points of the cornea at the highest concavity [highest concavity peak distance (HCPD)], central concave curvature at the highest concavity [highest concavity radius (HCR)]. To assess the change of parameters by disease severity, the keratoconus group was divided into two subgroups, and their biomechanical parameters were compared with each other and with normal group. The parameters' predictive ability was assessed by receiver operating characteristic (ROC) curves. To control the effect of central corneal thickness (CCT) difference between the two groups, two subgroups with similar CCT were selected, and the analyses were repeated. RESULTS Of the 10 parameters compared, the means of the 8 were significantly different between groups (P < 0.05). Means of the parameters did not show significant difference between keratoconus subgroups (P > 0.05). ROC curve analyses showed excellent distinguishing ability for A1T and HCR [area under the curve (AUC) > 0.9], and good distinguishing ability for A2T, A2V, and HCDA (0.9 > AUC > 0.7). A1T reading was able to correctly identify at least 93% of eyes with keratoconus (cut-off point 7.03). In two CCT matched subgroups, A1T showed an excellent distinguishing ability again. CONCLUSIONS The A1T seems a valuable parameter in the diagnosis of keratoconic eyes. It showed excellent diagnostic ability even when controlled for CCT. None of the parameters were reliable index for keratoconus staging.
منابع مشابه
Corneal Biomechanical Assessment Using Corneal Visualization Scheimpflug Technology in Keratoconic and Normal Eyes
Purpose. To compare the corneal biomechanical properties of keratoconic patients and age-matched controls using corneal visualization Scheimpflug technology (Corvis ST). Methods. Sixty keratoconic eyes from 47 keratoconus patients and 60 normal eyes from 60 controls were enrolled in this prospective study. Tomography and biomechanical parameters of all eyes were obtained with the Pentacam and C...
متن کاملEvaluation of Corneal Deformation Analyzed with Scheimpflug Based Device in Healthy Eyes and Diseased Ones
This study was designed to evaluate the correlation between corneal biomechanical and morphological data in healthy eyes, eyes that underwent myopic photorefractive keratectomy (PRK), keratoconus affected eyes, and keratoconus affected eyes that underwent corneal collagen crosslinking (CCC). Complete clinical eye examination of all eyes was followed by tomographic (Pentacam, Oculus, Wetzlar, Ge...
متن کاملFusion of Different Corneal Parameters to Improve the Diagnosis of Keratoconus
Purpose: To diagnose keratoconus from healthy eyes, as well as suspected keratoconus. Methods: Certain parameters were extracted from Casia, Corvis, and Pentacam HR devices for 3 groups of healthy, with keratoconus, and suspected keratoconus. This study was performed on 340 eyes with keratoconus, 310 normal eyes, and 350 suspected keratoconus. The processing method involved the fusion of featur...
متن کاملBiomechanical Characterization of Subclinical Keratoconus Without Topographic or Tomographic Abnormalities.
PURPOSE To present a case series of patients with subclinical keratoconus with normal topometric (anterior curvature) and tomographic findings in one eye who showed abnormalities detected by Corvis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany) in vivo biomechanical assessment. METHODS All patients had a complete ophthalmic examination, including the Corvis ST biomechanical measurements, opti...
متن کاملCorneal biomechanical data and biometric parameters measured with Scheimpflug-based devices on normal corneas.
AIM To analyze the correlations between ocular biomechanical and biometric data of the eye, measured by Scheimpflug-based devices on healthy subjects. METHODS Three consecutive measurements were carried out using the corneal visualization Scheimpflug technology (CorVis ST) device on healthy eyes and the 10 device-specific parameters were recorded. Pentacam HR-derived parameters (corneal curva...
متن کامل