Determination of the capacitance of solid-state potentiometric sensors: An electrochemical time-of-flight method.
نویسندگان
چکیده
A dual microelectrode electrochemical time-of-flight technique in which diffusion flux of Ag+, Cl-, or H+ ions electrochemically produced at a generator electrode is measured by recording potential-time transients with Ag, Ag/AgCl, or iridium oxide potentiometric microsensors, respectively, is developed. The generator and microsensor electrodes are typically spaced by 50-100 microm and are incorporated in the lithographically fabricated thin-layer-type devices. Under conditions of moderate rates of the ion electrogeneration, the potential-time (E-t) transients recorded with the three microsensors show excellent agreement with theory involving linear diffusion equations and the experimentally determined Nernstian slopes of the microsensors. However, when the generator current, or the initial concentration of the primary ion of interest is low, appreciable delays in the recorded E-t transients are observed due to the finite capacitance of the micropotentiometric sensors. The recorded delay in the E-t transients can be quantitatively accounted for by including the sensor capacitance (C) in the theoretical description of the transients. Direct comparison between the theoretical and the experimental E-t transients yields the sensor's capacitance. This capability of our new technique is unique in that it allows determination of the capacitance of a potentiometric sensor at open circuit. In the cases of silver electrodes, this method results in C = 31 +/- 2 microF/cm2, a value that is in agreement with those obtained by other methods. The results for silver chloride sensors yield a C in the range of 100-140 +/- 10 microF/cm2. The specific values depend on sensor preparation and the resulting roughness of the Ag/AgCl interface. Iridium oxide sensors show a capacitance that linearly depends on the thickness of the film. Specific capacitance of these microporous films was determined to be 59 +/- 6 F/cm3.
منابع مشابه
Measurements of the capacitance and the response time of solid-state potentiometric sensors by an electrochemical time-of-flight method
A micro-electrochemical time-of-flight technique relying on galvanostatic ion generation and potentiometric sensing (P-ETOF) has been developed in order to characterize dynamic properties of solid-state potentiometric sensors and to measure their capacitance. Lithographically-fabricated generator-sensor devices featuring 10 lm-wide micro-electrodes spaced by 20– 50 lm are used in either an open...
متن کاملGreen Potentiometric Method for Determination of Triprolidine Hydrochloride and Pseudoephedrine Hydrochloride in their Different Pharmaceutical Matrices using Liquid and Solid Contact Gold Electrodes
Factors affecting the performance of polyvinyl chloride based ion selective electrode such as using different cationic exchangers and ionophores as well as sensor fabrication procedures were recently the scope of study by several researchers. This work presents an environment friendly electro-analytical method for the determination of Triprolidine HCl (TRI) and Pseudoephedrine HCl (PSE). A comp...
متن کاملDevelopment of novel potentiometric sensors for determination of lidocaine hydrochloride in pharmaceutical preparations, serum and urine samples
This article is focused on the determination of lidocaine hydrochloride as a local anaesthetic drug. A potentiometric method based on modified screen-printed and modified carbon paste ion-selective electrodes was described for the determination of lidocaine hydrochloride in different pharmaceutical preparations. It was based on potentiometric titration of lidocaine hydrochloride using modified ...
متن کاملDevelopment of novel potentiometric sensors for determination of lidocaine hydrochloride in pharmaceutical preparations, serum and urine samples
This article is focused on the determination of lidocaine hydrochloride as a local anaesthetic drug. A potentiometric method based on modified screen-printed and modified carbon paste ion-selective electrodes was described for the determination of lidocaine hydrochloride in different pharmaceutical preparations. It was based on potentiometric titration of lidocaine hydrochloride using modified ...
متن کاملDetermination of Chromium(III) and Magnesium(II) Ions in Pharmacological and Real Water Samples using Potentiometric Sensors based on Chitosan Schiff base Derivative as Green and Sensitive Ionophore
In the study, novel and sensitive carbon paste electrodes (CPEs) developed for thepotentiometric measurement of Cr(III) and Mg(II) ions in pharmacological and water samples.CPEs as indicator electrodes were prepared from a mixture of four components, includinggraphite powder, paraffin oil, multi-walled carbon nanotubes (MWCNTs), and a greenionophore (Chitosan Schiff base...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Analytical chemistry
دوره 78 18 شماره
صفحات -
تاریخ انتشار 2006