Optimization of a Dried Residue Specimen Preparation Method for Quantifying Analytes in Plutonium Metal Using Wdxrf
نویسندگان
چکیده
A novel method for preparing thin films was investigated for quantifying gallium and iron in plutonium solutions using WDXRF. This technique was developed to eliminate the potential for radioactive liquid to leak into the spectrometer, decrease specimen preparation time, and minimize waste. Samples were cast in μL quantities onto Kapton, and a surfactant was added to disperse the solution uniformly across the Kapton. After drying the specimens, they were sealed in a cell for analysis. Results to date indicate the method can provide a relative precision of ~0.5% for gallium and ~2% for iron, which is more than sufficient for routine sample analyses.
منابع مشابه
Improvements in Xrf Specimen Preparation Using the Dried Residue Method: Gallium in Plutonium Metal
Preparing dry specimens from liquid samples for XRF analysis avoids introducing caustic or hazardous liquids into the instrument. Several modifications were made to a dried residue specimen preparation method for quantifying gallium in plutonium metal in order to improve the method accuracy and precision. Ion exchange chromatography was utilized to remove the plutonium prior to casting the drie...
متن کاملAccurate Quantification of Radioactive Materials by X-ray Fluorescence: Gallium in Plutonium Metal
Two XRF specimen preparation methods were investigated for quantifying gallium in plutonium metal. Gallium in plutonium was chosen here as an example for demonstrating the efficacy of wavelength dispersive XRF for quantifying radioactive materials. The steps necessary to handle such materials safely will also be discussed. Quantification of plutonium samples by a wellestablished aqueous specime...
متن کاملQUANTIFICATION OF GALLIUM IN DRIED RESIDUE SAMPLES BY XRF: AN IMPROVED SAMPLE PREPARATION METHOD FOR ANALYZING PLUTONIUM METAL, C.G. Worley, pp. 421-426
A novel XRF sample preparation method was investigated to quantify the concentration of gallium in plutonium metal and minimize the possibility of contaminating the instrument with radioactive material. To prepare homogenous specimens and to add an internal standard, the plutonium must first be dissolved. The currently established method then entails removing the plutonium by chromatography and...
متن کاملACCURATE QUANTIFICATION OF DRIED RESIDUE THIN FILMS USING X-RAY FLUORESCENCE, C.G. Worley, G.J. Havrilla, pp. 392-397
An XRF specimen preparation method was developed to quantify the concentration of gallium in plutonium metal while minimizing the risk of contaminating the instrument with radioactive material. To ensure that homogenous specimens are examined, plutonium is dissolved in dilute HCl and HNOJ prior to analysis. In the preliminary work here, non-radioactive aqueous gallium standards were prepared, a...
متن کاملRadioactive Sample Effects on Edxrf Spectra
Energy dispersive X-ray fluorescence (EDXRF) is a rapid, straightforward method to determine sample elemental composition. A spectrum can be collected in a few minutes or less, and elemental content can be determined easily if the detector has adequate energy resolution. Radioactive alpha emitters, however, emit X-rays during the alpha decay process that complicates spectral interpretation. Thi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -