Comparison of TLD calibration methods for 192Ir dosimetry
نویسندگان
چکیده
For the purpose of dose measurement using a high-dose rate (192)Ir source, four methods of thermoluminescent dosimeter (TLD) calibration were investigated. Three of the four calibration methods used the (192)Ir source. Dwell times were calculated to deliver 1 Gy to the TLDs irradiated either in air or water. Dwell time calculations were confirmed by direct measurement using an ionization chamber. The fourth method of calibration used 6 MV photons from a medical linear accelerator, and an energy correction factor was applied to account for the difference in sensitivity of the TLDs in (192)Ir and 6 MV. The results of the four TLD calibration methods are presented in terms of the results of a brachytherapy audit where seven Australian centers irradiated three sets of TLDs in a water phantom. The results were in agreement within estimated uncertainties when the TLDs were calibrated with the (192)Ir source. Calibrating TLDs in a phantom similar to that used for the audit proved to be the most practical method and provided the greatest confidence in measured dose. When calibrated using 6 MV photons, the TLD results were consistently higher than the (192)Ir-calibrated TLDs, suggesting this method does not fully correct for the response of the TLDs when irradiated in the audit phantom.
منابع مشابه
Improvement of Accuracy in Environmental Dosimetry by TLD Cards Using Three-dimensional Calibration Method
Background: The angular dependency of response for TLD cards may cause deviation from its true value on the results of environmental dosimetry, since TLDs may be exposed to radiation at different angles of incidence from the surrounding area.Objective: A 3D setting of TLD cards has been calibrated isotropically in a standard radiation field to evaluate the improvement of the accuracy of meas...
متن کاملCalibration of iridium-192 source by ionization chamber for high dose rate brachytherapy
Background: The effectiveness and safety of brachytherapy treatment is mainly concerned with the calibration of sources and their traceability to internationally accepted Standards. Secondary Standard Dosimetry Laboratory (SSDL) does not offer calibration of ionization chambers with gamma-ray spectrum of high dose rate source 192Ir . This work has been carried out to calibrate the high dose rat...
متن کاملImprovement of Accuracy in Environmental Dosimetry by TLD Cards Using Three-dimensional Calibration Method
INTRODUCTION The angular dependency of response for TLD cards may cause deviation from its true value on the results of environmental dosimetry, since TLDs may be exposed to radiation at different angles of incidence from the surrounding area. OBJECTIVE A 3D setting of TLD cards has been calibrated isotropically in a standard radiation field to evaluate the improvement of the accuracy of meas...
متن کاملEvaluation of Effective Sources in Uncertainty Measurements of Personal Dosimetry by a Harshaw TLD System
Background: The accurate results of the individual doses in personal dosimety which are reported by the service providers in personal dosimetry are very important. There are national / international criteria for acceptable dosimetry system performance.Objective: In this research, the sources of uncertainties are identified, measured and calculated in a personal dosimetry system by TLD.Method: T...
متن کاملAssessment of the effects of radiation type and energy on the calibration of TLD-100
Introduction: In radiation therapy, knowing the dose rates to healthy organs and tumors is beneficial, and thermoluminescent dosimeter (TLD) allows for this possibility. This study was aimed at determining the dose-response differences of TLDs in various types of radiation, energy levels, and dose rate calibrated with other types of radiation beams and energy and dose levels. Materials and Meth...
متن کامل