Perfused hypertonic-saline-augmented needle enlarges ablation zones in ex vivo porcine livers

نویسندگان

  • Caoye Wang
  • Qi Wang
  • Dachuan Zhang
  • Zhongming He
  • Wenhua Chen
چکیده

There is a great clinical requirement to improve radiofrequency ablation (RFA) efficacy and create larger coagulation necrotic areas. The aim of the present study was to assess the ability of a hypertonic-saline (HS)-enhanced multipolar RFA technique using a perfused electrode to increase RF-created coagulation necrosis, and to compare that technique with natural saline-augmented needle and conventional multipolar RFA. A total of 18 ablations were performed in explanted porcine livers. A total of 6 thermal ablation zones were created in each of 3 groups treated with the conventional multipolar mode, the multipolar mode with 0.9% NaCl and the multipolar mode with 6% NaCl, respectively. During RFA, the dimensions and volumes of the ablation zones were compared, and gross and microscopic pathological evaluations were performed. Multipolar RFA with 6% NaCl created the largest short-axis diameters and volumes of coagulation necrosis (3.89±0.09 mm and 40.01±2.86 mm3, respectively) among the three groups (conventional group: 2.31±0.04 mm and 8.99±0.52 mm3, respectively; 0.9% NaCl solution group: 3.17±0.05 mm and 21.79±1.05 mm3, respectively). Overall, multipolar RFA with the instillation of 6% NaCl solution through an open perfusion system created a larger ablation zone compared with the conventional and 0.9% NaCl modes. Therefore, HS-enhanced multipolar RFA may be a promising approach for treating large liver tumors.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Hepatic Radiofrequency Ablation Using Multiple Probes: Ex Vivo and In Vivo Comparative Studies of Monopolar versus Multipolar Modes

OBJECTIVE We wanted to compare the efficiency of multipolar radiofrequency ablation (RFA) using three perfused-cooled electrodes with multiple overlapping and simultaneous monopolar techniques for creating an ablation zone in ex vivo bovine livers and in in vivo porcine livers. MATERIALS AND METHODS In the ex vivo experiments, we used a 200 W generator (Valleylab, CC-3 model) and three perfus...

متن کامل

Optimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers

OBJECTIVE To determine the optimized protocol for wet monopolar radiofrequency ablation (RFA) using a perfused-cooled electrode to induce coagulation necrosis in the ex vivo bovine liver. MATERIALS AND METHODS Radiofrequency was applied to excised bovine livers in a monopolar mode using a 200W generator with an internally cooled electrode (groups A and B) or a perfused-cooled electrode (group...

متن کامل

Experimental Evaluation of the Heat Sink Effect in Hepatic Microwave Ablation

PURPOSE To demonstrate and quantify the heat sink effect in hepatic microwave ablation (MWA) in a standardized ex vivo model, and to analyze the influence of vessel distance and blood flow on lesion volume and shape. MATERIALS AND METHODS 108 ex vivo MWA procedures were performed in freshly harvested pig livers. Antennas were inserted parallel to non-perfused and perfused (700,1400 ml/min) gl...

متن کامل

A Comparative Experimental Study of the In-vitro Efficiency of Hypertonic Saline-Enhanced Hepatic Bipolar and Monopolar Radiofrequency Ablation

OBJECTIVE To compare the in-vitro efficiency of a hypertonic saline (HS)-enhanced bipolar radiofrequency (RF) system with monopolar RF applications by assessing the temperature profile and dimensions of RF-created coagulation necrosis in bovine liver. MATERIALS AND METHODS A total of 27 ablations were performed in explanted bovine livers. After placement of two 16-gauge open-perfused electrod...

متن کامل

Comparison of Wet Radiofrequency Ablation with Dry Radiofrequency Ablation and Radiofrequency Ablation Using Hypertonic Saline Preinjection: Ex Vivo Bovine Liver

OBJECTIVE We wished to compare the in-vitro efficiency of wet radiofrequency(RF) ablation with the efficiency of dry RF ablation and RF ablation with preinjection of NaCl solutions using excised bovine liver. MATERIALS AND METHODS Radiofrequency was applied to excised bovine livers in a monopolar mode for 10 minutes using a 200 W generator and a perfused-cooled electrode with or without injec...