Development and evaluation of a candidate reference method for the determination of total cortisol in human serum using isotope dilution liquid chromatography/mass spectrometry and liquid chromatography/tandem mass spectrometry.
نویسندگان
چکیده
Cortisol is an important diagnostic marker for the production of steroid hormones, and accurate measurements of serum cortisol are necessary for proper diagnosis of adrenal function. A candidate reference method involving isotope dilution coupled with liquid chromatography/mass spectrometry (LC/MS) and liquid chromatography/tandem mass spectrometry (LC/MS/MS) has been developed and critically evaluated. An isotopically labeled internal standard, cortisol-d(3), was added to serum, followed by equilibration and solid-phase and ethyl acetate extractions to prepare samples for liquid chromatography/mass spectrometry electrospray ionization (LC/MS-ESI) and liquid chromatography/tandem mass spectrometry electrospray ionization (LC/MS/MS-ESI) analyses. (M + H)(+) ions at m/z 363 and 366 for cortisol and its labeled internal standard were monitored for LC/MS. The transitions of (M + H)(+) --> [(M + H)(+) - 2H(2)O] at m/z 363 --> 327 and 366 --> 330 were monitored for LC/MS/MS. The accuracy of the measurement was evaluated by a comparison of results of this candidate reference method on lyophilized human serum reference materials for cortisol [Certified Reference Materials 192 and 193] with the certified values determined by gas chromatography/mass spectrometry reference methods and by a recovery study for the added cortisol. The results of this method for total cortisol agreed with the certified values within 1.1%. The recovery of the added cortisol ranged from 99.8% to 101.0%. This method was applied to the determination of cortisol in samples of frozen serum pools. Excellent precision was obtained with within-set CVs of 0.3%-1.5% and between-set CVs of 0.04%-0.4% for both LC/MS and LC/MS/MS analyses. The correlation coefficients of all linear regression lines ranged from 0.998 to 1.000. The detection limits (at a signal-to-noise ratio of approximately 3-5) were 10 and 15 pg for LC/MS and LC/MS/MS, respectively. This method, which demonstrates good accuracy and precision, and is free from interferences from structural analogues, qualifies as a candidate reference method and can be used as an alternative reference method to provide an accuracy base to which the routine methods can be compared.
منابع مشابه
Liquid Chromatography Tandem Mass/Mass Spectrometry for the Quantification of Fudosteine in Human Serum without Precolumn Derivatization
A quantitative analysis method for fudosteine in human serum by high-performance liquid chromatography-electrospray ionization tandem mass spectrometry (HPLC-ESI/MS/MS) was established, which shows high sensitivity and selectivity. The mobile phase composition was 75% 20 mM acetic acid and 25% acetonitril, which was pumped at a flow rate of 0.40 mL/min. The overall chromatographic run time was ...
متن کاملLiquid Chromatography Tandem Mass/Mass Spectrometry for the Quantification of Fudosteine in Human Serum without Precolumn Derivatization
A quantitative analysis method for fudosteine in human serum by high-performance liquid chromatography-electrospray ionization tandem mass spectrometry (HPLC-ESI/MS/MS) was established, which shows high sensitivity and selectivity. The mobile phase composition was 75% 20 mM acetic acid and 25% acetonitril, which was pumped at a flow rate of 0.40 mL/min. The overall chromatographic run time was ...
متن کاملDevelopment of a candidate reference method for determination of tyrosine in serum by isotope dilution liquid chromatography-tandem mass spectrometry.
An isotope dilution liquid chromatography/tandem mass spectrometry is proposed as a reference method to determine the level of tyrosine in human serum. The advantages of this method include simple sample preparation without derivatization, the selective detection of compounds of interest in complex matrices, and the use of an isotopically labeled analogue as an internal standard. Tyrosine and i...
متن کاملDevelopment and Validation of Bioanalytical Method for Simultaneous Estimation of Nebivolol Enantiomers in Human Plasma Using Liquid Chromatography-tandem Mass Spectrometry
The present study describes a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the simultaneous determination of S-RRR and R-SSS nebivolol (nebivolol enantiomers) in human plasma using solid phase extraction technique. Method of both S-RRR and R-SSS nebivolol (nebivolol enantiomers) has been developed and validated using racemic nebivolol D4 as an internal standard. Analytes...
متن کاملComparative Pharmacokinetics of Berberine, Palmatine and Jatrorrhizine in Rat Plasma after Oral Administration of Rhizoma coptidis and Zuojinwan Using Liquid Chromatography-Tandem Mass Spectrometry
A selective and sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed for the determination of berberine, palmatine and jatrorrhizine in rat plasma. Target compounds, together with the internal standard (metronidazole), were extracted from rat plasma samples by protein precipitation with acetonitrile-methanol (1:2, v/v). Chromatography was carried out using a ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Analytical chemistry
دوره 76 4 شماره
صفحات -
تاریخ انتشار 2004