An Accurate Metabolic Simulator for Indirect Calorimetry System Validation
نویسنده
چکیده
A metabolic simulator capable of simulation of respiratory oxygen consumption (VO2) and carbon dioxide production (VCO2) for validation of indirect calorimetry systems is described. The metabolic simulator adds/removes respiratory gases by way of mass flow controllers to/from a physiologically realistic lung simulator (Michigan Instruments Inc. Training/Test Lung). The metabolic simulator and Michigan Instruments Training/Test Lung in combination are capable of simulating a wide range of metabolic and physiologic lung conditions. Simulation of VCO2 is achieved by adding controlled amounts of carbon dioxide to the Training/Test Lung every (mechanical ventilator delivered) breath cycle. VO2 simulation consists of adding controlled amounts of nitrogen and removing controlled amounts of inspiratory gas from the Training/Test Lung every breath. Real time monitoring of ventilation parameters is obtained from a Novametrix Cosmo+ respiratory profile monitor and control of gas flow through the simulator is achieved by way of a programmed IBM personal computer.
منابع مشابه
New resting energy expenditure prediction equations for patients with rheumatoid arthritis.
OBJECTIVES Resting energy expenditure (REE), one of the main components of total energy expenditure, can be measured via indirect calorimetry and/or predicted from equations. The latter may be misleading in RA, as they do not take into account the metabolic alterations occurring in RA. The objectives of this study are to evaluate the accuracy of widely used REE-predictive equations in RA patien...
متن کاملReal-time estimation of daily physical activity intensity by a triaxial accelerometer and a gravity-removal classification algorithm.
We have recently developed a simple algorithm for the classification of household and locomotive activities using the ratio of unfiltered to filtered synthetic acceleration (gravity-removal physical activity classification algorithm, GRPACA) measured by a triaxial accelerometer. The purpose of the present study was to develop a new model for the immediate estimation of daily physical activity i...
متن کاملMetaCheck Accuracy Validation Report
The MetaCheck (Korr Medical Technologies, Salt Lake City, Utah) is an instrument designed to measure Resting Metabolic Rate (RMR) using indirect calorimetry. Indirect calorimetry is a method of calculating metabolic rate from the measured the amount of oxygen consumed by the body. Using the MetaCheck mouthpiece, the individual being tested breathes in room air and the gas the person breathes ou...
متن کاملEstimation of basal metabolic rate in Chinese: are the current prediction equations applicable?
BACKGROUND Measurement of basal metabolic rate (BMR) is suggested as a tool to estimate energy requirements. Therefore, BMR prediction equations have been developed in multiple populations because indirect calorimetry is not always feasible. However, there is a paucity of data on BMR measured in overweight and obese adults living in Asia and equations developed for this group of interest. The a...
متن کاملComparison of indirect calorimetry, the Fick method, and prediction equations in estimating the energy requirements of critically ill patients.
BACKGROUND Accurate measurement of resting energy expenditure (REE) is helpful in determining the energy needs of critically ill patients requiring nutritional support. Currently, the most accurate clinical tool used to measure REE is indirect calorimetry, which is expensive, requires trained personnel, and has significant error at higher inspired oxygen concentrations. OBJECTIVE The purpose ...
متن کامل