An in vitro study of dural lesions produced by 25-gauge Quincke and Whitacre needles evaluated by scanning electron microscopy.
نویسندگان
چکیده
BACKGROUND AND OBJECTIVES A study using scanning electron microscopy showed that although the laminas forming the dura mater are concentric and parallel to the surface of the medulla, the fiber layers' orientations are different in each sub-lamina, dispelling the conventional knowledge that all the fibers of the dura are arranged in a parallel direction. Thus, this study evaluated the dural lesions produced by Whitacre and Quincke spinal needles in the external and internal surface of the dura mater of the lower spine area in an attempt to gain more insight into the pathophysiology of postdural puncture headaches (PDPH). METHODS The T11-L4 dural membranes from 5 fresh (immediately after extraction of organs for transplantation), male patients declared brain dead, ages 23, 46, 48, 55, and 60 years, were excised by anterior laminectomy. Morphologic orientation of the membrane and normal pH were maintained with an apparatus designed for this purpose. One hundred punctures (20 on each sample) at 90-degree angles were done with a new needle each time, 50 with 25-gauge Whitacre and 50 with 25-gauge Quincke needles. Half of the punctures with the Quincke needles were done with the bevel in parallel direction to the axis of the spinal cord, and the rest with the bevel perpendicular to it. Fixation in solutions of 2.5% glutaraldehyde phosphate buffer, followed by dehydration with acetone, was done 15 minutes after the punctures. After acetone was removed at ideal conditions of temperature and pressure, the specimens were then metallized with carbon followed by gold and inspected under a scanning electron microscope. RESULTS Twenty-five of the Whitacre and 23 of the Quincke punctures were found for evaluation. There were no differences in the cross-sectional area of the punctures produced by the Whitacre or Quincke needles on the dura. The area of the dural lesions produced by 25-gauge Quincke needles, 15 minutes after they have been withdrawn, was 0.023 mm2 (confidence interval [CI] 95%, 0.015 to 0.027) in the external aspect (epidural surface) and 0.034 mm2 (CI 95%, 0.018 to 0.051) in the internal aspect (arachnoid surface) of the dural sac. The area of the lesions produced by the 25-gauge Whitacre needles was 0.026 mm2 (CI 95%, 0.019 to 0.032) and 0.030 mm2 (CI 95%, 0.025 to 0.036) in the external and internal surfaces of the dural sac, respectively. There were no significant differences in the cross-sectional areas of the punctures produced by the 25-gauge Whitacre or 25-gauge Quincke needles. Moreover, with Quincke needles the dural lesions closed in an 88.3% (CI 95%, 86.3 to 92.4) and 82.7% (CI 95%, 74.1 to 90.9) of their original sizes in the epidural and arachnoid surfaces, respectively. With Whitacre needles, the closure occurred in an 86.8% (CI 95%, 83.8 to 90.3) and 84.8% (CI 95% 81.7 to 87.3) in the dural and arachnoid surfaces, respectively. However, there were differences in the morphology of the lesions. The Whitacre needles produced coarse lesions with significant destruction in the dura's fibers while the Quincke needles produced a 'U'-shaped lesion (flap) that mimics the opened lid of a tin can, regardless of the tip's direction. CONCLUSIONS The needles produced lesions in the dura with different morphology and characteristics. Lesions with the Quincke needles resulted in a clean-cut opening in the dural membrane while the Whitacre needle produced a more traumatic opening with tearing and severe disruption of the collagen fibers. Thus, we hypothesized that the lower incidence of PDPH seen with the Whitacre needles may be explained, in part, by the inflammatory reaction produced by the tearing of the collagen fibers after dural penetration. This inflammatory reaction may result in a significant edema which may act as a plug limiting the leakage of cerebrospinal fluid.
منابع مشابه
Dura-arachnoid lesions produced by 22 gauge Quincke spinal needles during a lumbar puncture.
AIMS The dural and arachnoid hole caused by lumbar puncture needles is a determining factor in triggering headaches. The aim of this study is to assess the dimensions and morphological features of the dura mater and arachnoids when they are punctured by a 22 gauge Quincke needle having its bevel either in the parallel or in the transverse position. METHODS Fifty punctures were made with 22 ga...
متن کاملA Randomized Double Blind Study to Find the Incidence of Post Dural Puncture Headache Using Quincke and Whitacre Needles
The objectives of this study is to find the incidence of post dural puncture headache(PDPH) while using traumatic(Quincke) and atraumatic(Whitacre) needles for spinal anaesthesia . A prospective randomized double blind study in 100 ASA 1 and 2 patients undergoing lumbar subarachnoid block for elective inguinal hernia surgery conducted over 8 months. Lumbar puncture was performed by Anaesthesiol...
متن کاملThe effect of spinal needle design, size, and penetration angle on dural puncture cerebral spinal fluid loss.
Postdural puncture headache (PDPH) is a debilitating side effect of spinal anesthesia, the result of dural puncture and cerebrospinal fluid (CSF) leakage with an incidence of 3% to 75% in patient populations. Despite numerous in vitro and in viva studies that have identified predictors associated with PDPH, debate continues on the best technique to reduce CSF leak after dural puncture. The purp...
متن کاملPost dural puncture headache after spinal anaesthesia for caesarean section: a comparison of 25 g Quincke, 27 g Quincke and 27 g Whitacre spinal needles.
OBJECTIVE To compare the frequency and severity of post dural puncture headache in obstetric patients using 25G Quincke, 27G Quincke and 27G Whitacre spinal needles. DESIGN Comparative, randomized, double-blind, interventional study. PLACE AND DURATION OF STUDY Liaquat University Hospital Hyderabad from October 2005 to December 2006. PATIENTS AND METHODS 480 ASA I-II full term pregnant wo...
متن کاملComparison of Spinal Needle Deflection in a Ballistic Gel Model
BACKGROUND Percutaneous diagnostic and therapeutic procedures are commonly used in the treatment of spinal pain. The success of these procedures depends on the accuracy of needle placement, which is influenced by needle size and shape. OBJECTIVES The purpose of this study is to examine and quantify the deviation of commonly used spinal needles based on needle tip design and gauge, using a bal...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Regional anesthesia and pain medicine
دوره 25 4 شماره
صفحات -
تاریخ انتشار 2000