A routine quality assurance test for CT automatic exposure control systems

نویسندگان

  • Gareth R. Iball
  • Alexis C. Moore
  • Elizabeth J. Crawford
چکیده

The study purpose was to develop and validate a quality assurance test for CT automatic exposure control (AEC) systems based on a set of nested polymethylmethacrylate CTDI phantoms. The test phantom was created by offsetting the 16 cm head phantom within the 32 cm body annulus, thus creating a three part phantom. This was scanned at all acceptance, routine, and some nonroutine quality assurance visits over a period of 45 months, resulting in 115 separate AEC tests on scanners from four manufacturers. For each scan the longitudinal mA modulation pattern was generated and measurements of image noise were made in two annular regions of interest. The scanner displayed CTDIvol and DLP were also recorded. The impact of a range of AEC configurations on dose and image quality were assessed at acceptance testing. For systems that were tested more than once, the percentage of CTDIvol values exceeding 5%, 10%, and 15% deviation from baseline was 23.4%, 12.6%, and 8.1% respectively. Similarly, for the image noise data, deviations greater than 2%, 5%, and 10% from baseline were 26.5%, 5.9%, and 2%, respectively. The majority of CTDIvol and noise deviations greater than 15% and 5%, respectively, could be explained by incorrect phantom setup or protocol selection. Barring these results, CTDIvol deviations of greater than 15% from baseline were found in 0.9% of tests and noise deviations greater than 5% from baseline were found in 1% of tests. The phantom was shown to be sensitive to changes in AEC setup, including the use of 3D, longitudinal or rotational tube current modulation. This test methodology allows for continuing performance assessment of CT AEC systems, and we recommend that this test should become part of routine CT quality assurance programs. Tolerances of ± 15% for CTDIvol and ± 5% for image noise relative to baseline values should be used.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Estimation of Absorbed and Effective Doses in Organs through Computed Tomography Examinations Using Automatic Exposure Control and Fixed Tube Current Techniques: A Phantom Case Study

Introduction: The study aimed to assess absorbed and effective doses in organs through computed tomography (CT) examinations using automatic exposure control (AEC) and fixed tube current (FTC) techniques. Material and Methods: Scanning parameters were obtained for routine adult CT examinations and used to estimate the organ absorbed and effec...

متن کامل

A quality assurance program for an amorphous silicon electronic portal imaging device using in-house developed phantoms: a method development for dosimetry purposes

Background: Electronic portal imaging devices (EPIDs) play an important role in radiation therapy portal imaging, geometric and dosimetric verifications. A successful utilization of EPIDs for imaging and dosimetric purposes requires a reliable quality control process routine to be carried out regularly. In this study, two in-house phantoms were developed and analyzed for implementation in a qua...

متن کامل

Quality assurance for image-guided radiation therapy utilizing CT-based technologies: a report of the AAPM TG-179.

PURPOSE Commercial CT-based image-guided radiotherapy (IGRT) systems allow widespread management of geometric variations in patient setup and internal organ motion. This document provides consensus recommendations for quality assurance protocols that ensure patient safety and patient treatment fidelity for such systems. METHODS The AAPM TG-179 reviews clinical implementation and quality assur...

متن کامل

Patient dose in routine X-ray examinations in Yazd state

Background: Medical x-rays are the largest man-made source of public exposure to ionizing radiation. It is important to avoid conditions where the amount of radiation used is more than that needed for the procedure. Materials and Methods: The Entrance Skin Exposure (ESE) measurement was conducted for quality control of x-ray machines and survey of operator’s experimental techniques. The ESEs we...

متن کامل

Radiation protection principles observance in Iranian dental schools

Background: In recent decades many guidelines has been conducted by radiation protection organizations about radiation protection in dentistry. This study aimed to evaluate the observance of these guidelines in educational clinics of all dental schools in Iran. Material and Methods: In this cross-sectional study a self-administered questionnaire, based on National Radiation Protection ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016