Effect of digital noise reduction on the accuracy of endodontic file length determination
نویسندگان
چکیده
PURPOSE The aim of the present study was to evaluate the measurement accuracy of endodontic file length on periapical digital radiography after application of noise reduction digital enhancement. MATERIALS AND METHODS Thirty-five human single-rooted permanent teeth with canals measuring 20-24 mm in length were selected. ISO #08 endodontic files were placed in the root canals of the teeth. The file lengths were measured with a digital caliper as the standard value. Standard periapical digital images were obtained using the Digora digital radiographic system and a dental X-ray unit. In order to produce the enhanced images, the noise reduction option was applied. Two blinded radiologists measured the file lengths on the original and enhanced images. The measurements were compared by repeated measures ANOVA and the Bonferroni test (α=0.05). RESULTS Both the original and enhanced digital images provided significantly longer measurements compared with the standard value (P<0.05). There were no significant differences between the measurement accuracy of the original and enhanced images (P>0.05). CONCLUSION Noise reduction digital enhancement did not influence the measurement accuracy of the length of the thin endodontic files on the digital periapical radiographs despite the fact that noise reduction could result in the elimination of fine details of the images.
منابع مشابه
Accuracy of an Intraoral Digital System for File Length Measurement
Introdouction: Canal length measurement is essential for proper endodontic treatment. Any error or miscalculation in working length determination, particularly in curved canals, can result in complications during or after root canal therapy. Digital radiography has enabled accurate measurement of curved canal length. In this study, we evaluated the accuracy of calibration of a complementary m...
متن کاملIn Vitro Comparative Study of the Working Length Determination with Radiovisiography and Conventional Radiography in Dilacerated Canals
Introduction: The aim of this study was to compare digital and conventional radiography in determining the working length of dilacerated canals.Methods: Thirty nine human extracted single-rooted teeth with root curvature more than 35 degrees were included in this study. After access preparation, a file was inserted into the canal and advanced until the file tip was visualized at the foramen. Wi...
متن کاملبررسی عملکرد کالیبراسیون نرمافزار سیستم دیجیتال CMOS در ارزیابی طول کارکرد کانال
Background and Aims: Estimating the canal length is essential for reaching to a proper endodontic treatment. Any error and miscalculation in estimation of the working length, especially in curved canals, can result in complication during and after root canal therapy. The introduction of digital radiography has enabled us to measure curved canal length. The purpose of this study was to evaluat...
متن کاملمقایسه دقت رادیوگرافی دیجیتال با رادیوگرافی معمولی در ارزیابی طول کانالهای انحنادار
Background and Aim: Proper shaping and cleaning of the root canal system is one of the most important aspects of endodontic treatment. To estimate the canal length before instrumentation in endodontic treatment, traditionally, conventional radiographic techniques and recently Direct Digital Radiography (DDR) are applied. The application of computer technology to radiography has allowed less exp...
متن کاملIn vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...
متن کامل