Matrix effects during analysis of plasma samples by electrospray and atmospheric pressure chemical ionization mass spectrometry: practical approaches to their elimination.
نویسندگان
چکیده
Some cases of occurrence of matrix effects (mostly ion suppression) in protein-precipitated plasma samples, and their influence on the validity of plasma concentrations and pharmacokinetic parameters, are discussed. The comparison of matrix effects using either electrospray (TurboIonspray, TISP) or atmospheric pressure chemical ionization (APCI) indicated that APCI is less prone to matrix effects. Nevertheless, TISP is usually the first choice of ionization technique since unknown thermally labile metabolites might be present in the plasma samples causing erroneous results. A high impact of ion suppression on the plasma concentrations after intravenous (i.v.) administration was found, depending on the drug formulation (vehicle). Since ion suppression caused significantly lower plasma concentrations (by a factor of up to 5.5) after i.v. dosing, the area under the curve (AUC) was underestimated and the plasma clearance was consequently erroneously high, with an impact on drug candidate selection. By simple stepwise dilution (e.g. 10-fold and 50-fold) of the supernatant of protein-precipitated plasma samples, including all calibration and quality control samples, the matrix effects were recognized and eliminated.
منابع مشابه
Comparison of ESI– and APCI–LC–MS/MS methods: A case study of levonorgestrel in human plasma☆
Electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI) techniques for liquid chromatography-tandem mass spectrometry (LC-MS/MS) determination of levonorgestrel were evaluated. In consideration of difference in ionization mechanism, the two ionization sources were compared in terms of LC conditions, MS parameters and performance of method. The sensitivity for detection...
متن کاملDetecting Trace Explosives and Formulations Using Laser Electrospray Mass Spectrometry
Mass analysis using laser electrospray mass spectrometry (LEMS) is demonstrated for the detection of trace samples of explosives at atmospheric pressure directly from a substrate. A non-resonant femtosecond duration laser pulse vaporizes native samples at atmospheric pressure for subsequent electrospray ionization and transfer into a mass spectrometer. LEMS was used to detect < 1 µg of 2,3-dime...
متن کاملImaging of Proteins in Tissue Samples Using Nanospray Desorption Electrospray Ionization Mass Spectrometry.
Chemical maps of tissue samples provide important information on biological processes therein. Recently, advances in tissue imaging have been achieved using ambient ionization techniques, such as desorption electrospray ionization mass spectrometry (DESI-MS), but such techniques have been almost exclusively confined to the mapping of lipids and metabolites. We report here the use of nanospray d...
متن کاملOnline monitoring of chemical reactions by contactless atmospheric pressure ionization mass spectrometry.
Online monitoring of the appearance of the intermediates and products during chemical reactions is helpful to understand reaction mechanisms. Various analytical tools including nuclear magnetic resonance spectroscopy, Raman spectroscopy and mass spectrometry (MS) have been employed for the study. MS is one of the most promising means among analytical tools that can provide information of mass-t...
متن کاملMatrix effect in bio-analysis of illicit drugs with LC-MS/MS: influence of ionization type, sample preparation, and biofluid.
The purpose of the present work was to evaluate the synergistic effect of ionization type, sample preparation technique, and bio-fluid on the presence of matrix effect in quantitative liquid chromatography (LC)-MS/MS analysis of illicit drugs by post-column infusion experiments with morphine (10-microg/mL solution). Three bio-fluids (urine, oral fluid, and plasma) were pretreated with four samp...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Rapid communications in mass spectrometry : RCM
دوره 17 17 شماره
صفحات -
تاریخ انتشار 2003