Six of eight hemoglobin A1c point-of-care instruments do not meet the general accepted analytical performance criteria.

نویسندگان

  • Erna Lenters-Westra
  • Robbert J Slingerland
چکیده

BACKGROUND Hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments are widely used to provide rapid-turnaround results in diabetic care centers. We investigated the conformance of various Hb A(1c) POC instruments (In2it from Bio-Rad, DCA Vantage from Siemens, Afinion and Nycocard from Axis-Shield, Clover from Infopia, InnovaStar from DiaSys, A1CNow from Bayer, and Quo-Test from Quotient Diagnostics) with generally accepted performance criteria for Hb A(1c). METHODS The CLSI protocols EP-10, EP-5, and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using 3 certified secondary reference measurement procedures and the mean of the 3 reference methods. Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria, as calculated from analyses with 2 different reagent lot numbers for each Hb A(1c) method, was also evaluated. RESULTS Because of disappointing EP-10 results, 2 of the 8 manufacturers decided not to continue the evaluation. The total CVs from EP-5 evaluations for the different instruments with a low and high Hb A(1c) value were: In2it 4.9% and 3.3%, DCA Vantage 1.8% and 3.7%, Clover 4.0% and 3.5%, InnovaStar 3.2% and 3.9%, Nycocard 4.8% and 5.2%, and Afinion 2.4% and 1.8%. Only the Afinion and the DCA Vantage passed the NGSP criteria with 2 different reagent lot numbers. CONCLUSIONS Only the Afinion and the DCA Vantage met the acceptance criteria of having a total CV <3% in the clinically relevant range. The EP-9 results and the calculations of the NGSP certification showed significant differences in analytical performance between different reagent lot numbers for all Hb A(1c) POC instruments.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Three of 7 hemoglobin A1c point-of-care instruments do not meet generally accepted analytical performance criteria.

BACKGROUND In 2009, we investigated the conformance of 8 hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments. Since then, instruments have improved and new devices are available on the market. In this second study, we evaluated the performance of DCA Vantage, Afinion, InnovaStar, Quo-Lab, Quo-Test, Cobas B101, and B-analyst Hb A(1c) POC instruments. METHODS Clinical and Laboratory Sta...

متن کامل

Diagnosing diabetes mellitus: performance of hemoglobin A1c point-of-care instruments in general practice offices.

BACKGROUND Hemoglobin A1c (Hb A1c) measurement by hospital laboratory instruments, but not by point-of-care (POC) instruments, has been recommended for use to diagnose diabetes mellitus. We evaluated results from 13 Hb A1c external quality assurance (EQA) surveys over a 6-year period in Norway, from both POC instruments used in general practice (GP) offices and instruments in hospital laborator...

متن کامل

بررسی شاخص های مرتبط با کیفیت مراقبت بیماران دیابتی تحت پوشش برنامه پزشک خانواده روستایی شهرستان ارومیه در سال 1396

Background & Aims: More than 382 million people in the world live with diabetes, 80% of whom live in low-income and developing countries. Today, it is quite clear that improving metabolic control in diabetic patients can reduce the complications of the disease. The purpose of this study was to evaluate the quality of care associated with diabetic patients under the coverage of Urmia rural famil...

متن کامل

Few point-of-care hemoglobin A1c assay methods meet clinical needs.

Accurate and timely measurements of hemoglobin A1c (Hb A1c) 2 are an increasingly important challenge for clinical laboratories. As Hb A1c measurements are being used to answer new clinical questions, each new use of Hb A1c imposes requirements for the analytical performance of the methods used to measure it. Until recently, Hb A1c was used only as a key test for monitoring glycemic control in ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 56 1  شماره 

صفحات  -

تاریخ انتشار 2010