Static imaging of the electrical impedance tomography on cylinder physical phantom.
نویسندگان
چکیده
Static imaging of the electrical impedance tomography can obtain the absolute electrical conductivity distribution at one section of the subject. The test is performed on a cylinder physical phantom in which slim rectangle, hollow cylinder, small rectangle or three cylinders are selected to simulate complex conductivity perturbation objects. The measurement data is obtained by a data acquisition system with 32 compound electrodes. A group of static images of conductivity distribution in the cylinder phantom are reconstructed by the modified Newton-Raphson algorithm with two kinds of regularization methods. The results show correct position, size, conductivity difference, and similar shape of the perturbation objects in the images.
منابع مشابه
Performance improvements in a MF-EIT system for acute stroke: The UCL Mk2.5
A recently developed EITS system, the UCLH mk2.5 was tested on a resistor phantom and in linear time and frequency difference imaging of a cylindrical tank, 10cm diameter, filled with 0.1% saline, and 16 electrodes in a ring; the test object was a cylinder of banana, 2 cm long and 1 cm in diameter. Reproducible images with the same location as the test object were obtained with both methods. No...
متن کاملApplications of Electrical Impedance Tomography in Neurology
Introduction: Electrical impedance tomography (EIT) is a non-invasive technique utilized in various medical applications, including brain imaging and other neurological diseases. Recognizing the physiological and anatomical characteristics of organs based on their electrical properties is one of the main applications of EIT, as each variety of tissue structure has its own electrical characteris...
متن کاملEvaluation of the use of gold nanoparticle as contrast agent in electrical impedance imaging
Introduction: Electrical Impedance Tomography (EIT) is a real time and minimal invasive imaging modality for detection of lesions in tissues even when their structures show no changes. Early detection of lesion with EIT depends on difference between electrical impedance of targeted tissue and its surroundings. Therefore, finding a contrast agent for EIT that increase this diffe...
متن کاملDesign of Voltage Measuring Circuits in Electrical Impedance Tomography Based on Dual Frequency Phase-Sensitive Demodulator
In this paper, a dual frequency phase sensitive demodulation (PSD) technique is suggested to measure electrode voltages in electrical impedance tomography. The simulation results obtained by the dual frequency PSD are compared with the single frequency PSD in terms of accuracy on a simple phantom.
متن کاملElectrical conductivity imaging by magnetic resonance electrical impedance tomography (MREIT).
Magnetic resonance electrical impedance tomography (MREIT) is a recently developed imaging technique that combines MRI and electrical impedance tomography (EIT). In MREIT, cross-sectional electrical conductivity images are reconstructed from the internal magnetic field density data produced inside an electrically conducting subject when an electrical current is injected into the subject. In thi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Bio-medical materials and engineering
دوره 26 Suppl 1 شماره
صفحات -
تاریخ انتشار 2015