QA of intensity-modulated beams using dynamic MLC log files
نویسندگان
چکیده
To evaluate the utility of Dynalog file information for planar dose verification in IMRT QA, a program is developed to convert Dynalog file data to DMLC field files. For this study, five predefined fluencies are planned and delivered using Varian, Eclipse 3D planning system and 6MV photon beam of Varian, Clinac DMX linear accelerator. To measure planar dose distribution, Kodak, EDR2 films are exposed in similar setup as planning setup. Dynalog files are recorded for each delivery and converted into DMLC field files using in-house program. Delivered dose distributions are calculated using DMLC field files from Dynalog files. Planned, Measured and Delivered dose distributions are compared using gamma evaluation in Scanditronix, Omni Pro IMRT software. The Planned and Delivered planar dose distributions agree within 2% dose difference and 2 mm DTA. Measured dose distributions agree within 4% dose difference and 4 mm DTA with Planned dose distribution. Our results show Dynalog file as a promising tool for dynamic IMRT QA.
منابع مشابه
Inverse Verification of the Dose Distribution for Intensity Modulated Radiation Therapy Patient-specific Quality Assurance Using Dynamic MLC Log Files
The aim of this study was to investigate a novel method for verification of the dose distribution for intensity modulated radiation therapy (IMRT) patient-specific quality assurance (QA) using dynamic multi-leaf collimator (DMLC) log files (Dynalog files). Dynalog files are recorded every 50 ms by using a MLC controller during the IMRT treatment. Dynalog files contain actual MLC positional info...
متن کاملPractical approach for pretreatment verification of IMRT with flattening filter‐free (FFF) beams using Varian portal dosimetry
Patient-specific pretreatment verification of intensity-modulated radiation therapy (IMRT) or volumetric-modulated arc therapy (VMAT) is strongly recommended for all patients in order to detect any potential errors in treatment planning process and machine deliverability, and is thus performed routinely in many clinics. Portal dosimetry is an effective method for this purpose because of its pro...
متن کاملValidation of dynamic MLC-controller log files using a two-dimensional diode array.
Intensity-modulated radiation therapy (IMRT) delivered with multi-leaf collimator (MLC) in the step-and-shoot mode uses multiple static MLC segments to achieve intensity modulation. For typical IMRT treatment plans, significant numbers of segments are delivered with monitor units (MUs) of much less than 10. Verification of the ability of the linear accelerator (linac) to deliver small MU segmen...
متن کاملVerification of dynamic and segmental IMRT delivery by dynamic log file analysis
A program has been developed to evaluate the delivered fluence of step-and-shoot segmental and sliding window dynamic multileaf collimator (MLC) fields. To automate these checks, a number of tools have been developed using data available from the dynamic log files that can be created each time a dynamic delivery occurs. Experiments were performed with a Varian 2100EX with a 120 leaf MLC equippe...
متن کاملInvestigation of error detection capabilities of phantom, EPID and MLC log file based IMRT QA methods
A patient specific quality assurance (QA) should detect errors that originate anywhere in the treatment planning process. However, the increasing complexity of treatment plans has increased the need for improvements in the accuracy of the patient specific pretreatment verification process. This has led to the utilization of higher resolution QA methods such as the electronic portal imaging devi...