Multiplex Cytokine Profiling of Stimulated Mouse Splenocytes Using a Cytometric Bead-based Immunoassay Platform
نویسندگان
چکیده
Bead-based immunoassays employ the same basic principle as sandwich immunoassays. Capture beads, which can be differentiated by size and internal allophycocyanin (APC) fluorescence intensity, are conjugated to antibodies specific to a particular analyte. Next, a selected panel of defined capture bead sets is incubated with a biological sample containing target analytes specific to the capture antibodies. A biotinylated detection antibody cocktail is added, which leads to the formation of capture bead-analyte-detection antibody sandwiches. Finally, streptavidin-phycoerythrin (SA-PE) is added, which binds to biotinylated detection antibodies, providing fluorescent signal intensities in proportion to the amount of bound analyte. The PE fluorescent signal of analyte-specific beads regions is quantified using flow cytometry, and the concentrations of particular analytes are determined using data analysis software and the standard curve generated in the assay. In this experiment, we use a mouse T helper cytokine panel to simultaneously quantify the concentration of 13 separate cytokine targets in tissue culture supernatants collected from mouse splenocytes cultured under various stimulatory conditions.
منابع مشابه
An Evaluation of Commercial Fluorescent Bead-Based Luminex Cytokine Assays
The recent introduction of fluorescent bead-based technology, allowing the measurement of multiples analytes in a single 25-50 microl sample has revolutionized the study of cytokine responses. However, such multiplex approaches may compromise the ability of these assays to accurately measure actual cytokine levels. This study evaluates the performance of three commercially available multiplex c...
متن کاملBead-based multianalyte flow immunoassays: the cytometric bead array system.
Analytical cytometry has significant potential beyond cellular analysis. The inherent capability of flow cytometers to efficiently discriminate between uniformly sized particles based on their intrinsic properties provides the foundation for multiplex bead assays. The technology can be exploited in designing immunoassays, Western blot-like antibody assays, and nucleic acid hybridization assays....
متن کاملA novel bead-based fluorescence immunoassay for aldosterone
Aldosterone quantification helps evaluate the rennin-angiotensin-aldosterone system. The new bead-based multiplex platform has not been applied in aldosterone detection to achieve simultaneous measurements of multiple hormones. A new sensitive competitive bead immunoassay based on Luminex technology for detecting aldosterone in small sample volumes was developed using two-antibody coupled beads...
متن کاملDevelopment of a bead-based multiplex assay for simultaneous quantification of cytokines in horses.
The detection and quantification of equine cytokines has been hampered by the lack of antibodies for many years. With the development of antibody pairs for equine cytokines during the past years, the quantification of these essential regulators of the immune response became possible. After being successfully tested by enzyme-linked immunosorbent assays (ELISA), three of these anti-cytokine reag...
متن کاملEffects of Trauma-Hemorrhage and IL-6 Deficiency on Splenic Immune Function in a Murine Trauma Model
Splenic immune function is known to be depressed following hemorrhage. The present study investigates the effects of femoral shaft fracture, isolated or in combination with hemorrhage, on early stage cytokine production capacity of splenocytes and observes the role of IL-6 under these conditions. Male IL-6 knockout (IL-6(-/-)) and wild-type mice (WT) were randomly divided into three groups: sha...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 25 شماره
صفحات -
تاریخ انتشار 2017