Continuous blood pressure measurement using the pulse transit time: Comparison to intra-arterial measurement.

نویسندگان

  • Andreas Patzak
  • Yuri Mendoza
  • Heiko Gesche
  • Martin Konermann
چکیده

Continuous blood pressure (BP) measurement allows the investigation of transient changes in BP and thus may give insights into mechanisms of BP control. We validated a continuous, non-invasive BP measurement based on the pulse transit time (PTT), i.e., BP(PTT), by comparing it with the intra-arterial BP (BP(i.a.)) measurement. Twelve subjects (five females and seven males) were included. BP(i.a.) was obtained from the radial artery using a system from ReCor Medical. Systolic and diastolic BP were calculated using the PTT (BP(PTT), SOMNOscreen). (PTT) was determined from the electrocardiogram and the peripheral pulse wave. The BP was modulated by application of increasing doses of dobutamine (5, 10, 20 μg/kg body mass). Systolic BP(PTT) and systolic BP(i.a.) correlated significantly (R = 0.94). The limits of agreement in the Bland-Altman plot were ± 19 mmHg; the mean values differed by 1 mmHg. The correlation coefficient for the diastolic BP measurements was R = 0.42. The limits of agreement in the Bland-Altman plot were ± 18 mmHg, with a mean difference of 5 mmHg in favour of the BP(PTT). The study demonstrates a significant correlation between the measurement methods for systolic BP. The results encourage the application of PTT-based BP measurement for the evaluation of BP dynamics and pathological BP changes.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Designing and Constructing an Optical System to measure Continuous and Cuffless Blood Pressure Using Two Pulse Signals

Introduction Blood pressure (BP) is one of the important vital signs that need to be monitored for personal healthcare. Arterial blood pressure is estimated from pulse transit time (PTT). This study uses two pulse sensors to get PTT. The aim of this study was to construct an optical system and to monitor blood pressure continuously and without cuff in people with different ages. Materials and M...

متن کامل

A System for Continuous Estimating and Monitoring Cardiac Output via Arterial Waveform Analysis

Background: Cardiac output (CO) is the total volume of blood pumped by the heart per minute and is a function of heart rate and stroke volume. CO is one of the most important parameters for monitoring cardiac function, estimating global oxygen delivery and understanding the causes of high blood pressure. Hence, measuring CO has always been a matter of interest to researchers and clinicians. Sev...

متن کامل

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

In this paper, algorithm estimating the blood pressure was proposed using the pulse transit time (PTT) as a more convenient method of measuring the blood pressure. After measuring ECG and pressure pulse, and photoplethysmography, the PTT was calculated from the acquired signals. Thereafter, the system to indirectly measure the systolic pressure and the diastolic pressure was composed using the ...

متن کامل

Comparison of direct and indirect blood pressure in patients under esophagectomy surgery: a prospective observational study

Background: Arterial pressure is one of the most important physiological variables and often needs to be monitored repeatedly or continuously in perioperative period. Arterial pressure monitoring is one of the standard monitoring in operating room. During general anesthesia, blood pressure can be measured by using a noninvasive arterial pressure method or continuous invasive arterial pressure b...

متن کامل

Continuous and Cuffless Blood Pressure Monitoring Based on ECG and SpO2 Signals By Using Microsoft Visual C Sharp

Background: One of the main problems especially in operating room and monitoring devices is measurement of Blood Pressure (BP) by sphygmomanometer cuff.Objective: In this study we designed a new method to measure BP changes continuously for detecting information between cuff inflation times by using vital signals in monitoring devices. This will be achieved by extraction of the time difference ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Blood pressure

دوره 24 4  شماره 

صفحات  -

تاریخ انتشار 2015