A Novel Thin NIPAM Gel Cassette Dosimeter for Photon-Beam Radiotherapy
نویسندگان
چکیده
The response of thin polymer gel cassettes (called NIPAM gels) to ionizing radiation was investigated in this study. The NIPAM gels were prepared from gelatin, N-isopropyl acrylamide, tetrakis (hydroxymethyl) phosphoniumchloride, and N,N'-methylene-bis-acrylamide. Gel cassettes were irradiated in a phantom using a linear accelerator, and the polymerization morphology of irradiated NIPAM gel was characterized using scanning electron microscopy. The dose-response sensitivity of the NIPAM gels was evaluated using the differences in optical densities. The optical densities were obtained using a computer-controlled CCD camera that was connected to a planar illumination source for acquisition of optical transmission images. The central axis depth dose profiles of the phantom were extracted, and a comparison with ionization chamber measurements demonstrated similarities in profiles. The sensitivity, linearity of the response, accuracy, and reproducibility of the polymer gel cassettes were acceptable. However, the profiles of the half-blocked field irradiation showed no significant dispersion in the visible region. This study also extensively investigated the spatial stability of the NIPAM gel. The results showed that the gel cassette response remains stable for up to three months after irradiation.
منابع مشابه
Small field size dose-profile measurements: A comparison between Eclipse™ treatment planning system with NIPAM polymer gel dosimeter dose measurement
Introduction: Three physical conditions caused an external beam designated as a small field. The first is the loss of lateral charged particle equilibrium. The second is the effect of collimators on the primary photon source occlusion. The third condition is the size of the detector which is large compared to the field size. The most obvious perturbation effects in last condit...
متن کاملEvaluation of Gold Nanoparticle Size Effect on Dose Enhancement Factor in Megavoltage Beam Radiotherapy Using MAGICA Polymer Gel Dosimeter
Background: Gold nanoparticles (GNPs) are among the most promising radiosensitive materials in radiotherapy. Studying the effective sensitizing factors such as nanoparticle size, concentration, surface features, radiation energy and cell type can help to optimize the effect and possible clinical application of GNPs in radiation therapy. In this study, the radiation sensitive polymer gel was use...
متن کاملOptical and NMR dose response of N-isopropylacrylamide normoxic polymer gel for radiation therapy dosimetry.
BACKGROUND Application of less toxic normoxic polymer gel of N-isopropyl acrylamide (NIPAM) for radiation therapy has been studied in recent years. AIM In the current study the optical and NMR properties of NIPAM were studied for radiation therapy dosimetry application. MATERIALS AND METHODS NIPAM normoxic polymer gel was prepared and irradiated by 9 MV photon beam of a medical linac. The o...
متن کاملDose integration and dose rate characteristics of a NiPAM polymer gel MRI dosimeter system
The normoxic polymer gel dosimeter based on N-isopropyl acrylamide (NiPAM) is a promising full 3D-dosimeter with high spatial resolution and near tissue equivalency. NiPAM gel samples were irradiated to different doses using a linear accelerator. The absorbed dose was evaluated using MRI and statistical significance of the analysed data was calculated. The analysis was carried out using an in-h...
متن کاملThe basic radiation properties of the N-isopropylacrylamide based polymer gel dosimeter
Background: In this study the basic radiation properties of N-isopropylacrylamaide polymer gel dosimeter were determined together with verification of its soft tissue equivalency. Materials and Methods: The NIPAM gel was prepared and irradiated approximately 2 h after manufacturing. The magnetic resonance (MR) images were made 24 h after irradiation. The nuclear magnetic resonance (NMR) respons...
متن کامل