Digitizing Radiographic Films: a Simple Way to Evaluate Indirect Digital Images
نویسندگان
چکیده
OBJECTIVES This study applied a simple method to evaluate the performance of three digital devices (two scanners and one digital camera) using the reproducibility of pixel values attributed to the same radiographic image. METHODS Using the same capture parameters, a radiographic image was repeatedly digitized in order to determine the variability of pixel values given to the image throughout the digitization process. One coefficient value was obtained and was called pixel value reproducibility. RESULTS A significant difference in pixel values was observed among the three devices for the digitized images (ANOVA, p<0.00001). There was significant pixel value variability at the same digitization conditions for one scanner and the digital camera. CONCLUSIONS Digital devices may assign pixel values differently in consecutive digitization depending on the optical density of the radiographic image and the equipment. The pixel value reproducibility was not satisfactory as tested for two devices. It is maybe advisable knowing the digitization variations regarding pixel values whenever using digital radiography images in longitudinal clinical examinations.
منابع مشابه
The perceptibility curve test applied to CCD and two methods of digitization of dental film-based radiographs
Objectives: Several methods of image acquisition are accessible in dentistry. There is no overall acceptable method for image digitization so all different types of images can be comparable. The objective of this study was to compare the diagnostic accuracy of different methods of image digitization. Methods: This accuracy diagnostic test study used perceptibility curve test which first intr...
متن کاملEvaluation of the effect of four current Iranian processing solutions on the image quality of two dental X-ray films
Evaluation of the effect of four current Iranian processing solutions on the image quality of two dental X-ray films Dr. M. Imani Moghadam* - Dr. AHA. Afzali** - Eng. S. Ebrahim Zadeh*** *- Assistant Professor of Dentomaxillofacial Radiology Dept. - Faculty of Dentistry – Mashhad University of Medical Sciences. **- Radiologist. ***- The member of Mashhad University of Medical Sciences. Backgrou...
متن کاملارزیابی دقت دانسیتومتری تصاویر دیجیتال غیر مستقیم
Statement of Problem: One of the major goals, in most dental researches, is to measure bone destruction or deposition due to the progression or regression of disease. Failure of human eyes to detect minor radiographic density changes resulted in more accurate methods such as optical densitometry and direct or indirect digital densitometry.Purpose: The aim of this study was to determine the accu...
متن کاملمقایسه دقت فیلمهای Primax و Insight و تصاویر دیجیتالی CMOS-APS در تشخیص پوسیدگیهای جانبی دندانهای خلفی به صورت آزمایشگاهی
Introduction: Radiography is one of the most important diagnostic methods for evaluation of dental caries. On the other hand, the least amount of radiation along with high quality is a main gold standard. The purpose of this study was to compare the diagnostic accuracy of Primax, Insight Radiographic films, and direct digital images (Schick CMOS-APS), in detection of interproximal natural Carie...
متن کاملA Proposed Simple Teleradiology System Using a Digital Camera
Background & Aims: Teleradiology as an efficient way for avoiding unnecessary transfer of patients to special medical centers requires the images with digital format. However, in some developing countries, including Iran, imaging is being done using non digital devices. In such cases, changing analog images into digital ones can solve the problem. The method which is nowadays used for digita...
متن کامل