Introduction Wedge modifiers are commonly applied in external beam radiotherapy to change the dose distribution corresponding to the body contour and to obtain a uniform dose distribution within the target volume. Since the radiation dose delivered to the target must be within ±5% of the prescribed dose, accurate dose calculation by a treatment planning system (TPS) is important. The objective of the present study was to quantify the dose calculation accuracy of TiGRT TPS for physical wedged fields in radiotherapy. Materials and Methods A SemiflexTM ionization chamber was used for dose measurements in a water phantom; TiGRT TPS was also applied for dose calculations. The central axis (i.e., high dose-small dose gradient), build-up (i.e., high doselarge dose gradient), off-axis (i.e., high dose-small dose gradient), and out-of-field (i.e., low dose-small dose gradient) regions were evaluated in this study. Finally, the confidence limit values were obtained to quantify the dose calculation accuracy of TPS in these regions. Results The confidence limit values for the central axis, build-up, off-axis, and out-of-field regions were 1.01, 8.62, 1.79, and 55.24, respectively. Furthermore, the results showed that TiGRT TPS underestimated the dose of build-up and out-of-field regions for most points. Conclusion According to the results of the present study, it can be concluded that the dose calculation accuracy of TiGRT TPS for physical wedged fields in the central axis, build-up, and off-axis regions is adequate, while it is insufficient for out-of-field regions.
منابع مشابه
Assessment of Dose Calculation Accuracy of TiGRT Treatment Planning System for Physical Wedged fields in Radiotherapy
Introduction Wedge modifiers are commonly applied in external beam radiotherapy to change the dose distribution corresponding to the body contour and to obtain a uniform dose distribution within the target volume. Since the radiation dose delivered to the target must be within ±5% of the prescribed dose, accurate dose calculation by a treatment planning system (TPS) is important. The objective ...
متن کاملEvaluation of Three-dimensional Treatment Planning System (TPS) performance in dose calculation of virtual wedged fields using film dosimetry
Introduction: Nowadays radiotherapy plays an important role in cancer treatment. Different radiotherapy techniques improvement emphasizes on using of the precise ، appropriate and useful algorithms. one of these techniques are wedged which is used in radiotherapy to compensate missing tissues and create a uniform dose distribution in tissues. The Siemens Artiste linear accelera...
متن کاملDose Calculations for Lung Inhomogeneity in High-Energy Photon Beams and Small Beamlets: A Comparison between XiO and TiGRT Treatment Planning Systems and MCNPX Monte Carlo Code
Introduction Radiotherapy with small fields is used widely in newly developed techniques. Additionally, dose calculation accuracy of treatment planning systems in small fields plays a crucial role in treatment outcome. In the present study, dose calculation accuracy of two commercial treatment planning systems was evaluated against Monte Carlo method. Materials and Methods Siemens Once or linea...
متن کاملDose calculations accuracy of TiGRT treatment planning system for small IMRT beamlets in heterogeneous lung phantom
Background: Accurate dose calculations in small beamlets and lung material have been a great challenge for most of treatment planning systems (TPS). In the current study, the dose calculation accuracy of TiGRT TPS was evaluated for small beamlets in water and lung phantom by comparison to Monte Carlo (MC) calculations. Materials and Methods: The head of Siemens Oncor-impression linac...
متن کاملEvaluation of Accuracy and Quality assurance of external beam therapy with photons
Introduction: Receiving exact dose by the patients is vital in radiotherapy. In radiation therapy, the dosimetry of radiations is too important because of successful radiation inquires for delivering the exact dose to the target volume. This study is to evaluate the tolerances and the accuracy of calculated dose of photon beams in the treatment software system. The TECDOC1583 p...
متن کامل