Direct identification of bacteria from positive blood cultures by amplification and sequencing of the 16S rRNA gene: evaluation of BACTEC 9240 instrument true-positive and false-positive results.
نویسندگان
چکیده
In a previous study which evaluated the BACTEC 9240 automated blood culture system (Becton Dickinson Diagnostic Instrument Systems, Sparks, Md.), we noted a 1.3% "instrument false-positive" rate. That is, the BACTEC system signaled that a bottle (BACTEC Plus Aerobic/F bottle or BACTEC Anaerobic Lytic/10 bottle) culture was positive but a Gram stain was negative and there was no growth of bacteria or yeasts on subculture to chocolate agar. Furthermore, from the same sample of blood, cultures for fungi using the Isolator blood culture system (Wampole Laboratories, Cranbury, N.J.) were negative for growth. For the present study, we evaluated 76 instrument false-positive samples for the presence of 16S ribosomal DNA using the MicroSeq 500 kit (PE Biosystems, Foster City, Calif.). These samples also were negative for fungi by the Isolator method. This kit has a PCR module and sequencing module for the amplification and sequencing of the 16S RNA gene and provides a database for sequence alignment and identification of bacteria. To optimize the assay, we evaluated the effect of adding 0.5% bovine serum albumin to the sample from blood culture bottles and found that it decreased the effects of inhibitors on the PCR. Two control groups of blood culture specimens were also evaluated. One group (n = 45) were "instrument true positives"; the instrument signaled positive, and subsequent Gram stains were positive and subcultures on chocolate agar grew bacteria. The other group (n = 20) were "instrument true negatives"; the instrument signaled negative, the Gram stain was negative, and subcultures on chocolate agar and from the Isolator tube on fungal media showed no growth. None of the 76 instrument false-positive samples had evidence for 16S rRNA gene sequences. All of the instrument true-positive samples and all of the instrument true-negative specimens were positive and negative, respectively, using the MicroSeq 500 kit. Total peripheral white blood cell counts were statistically significantly higher for patients who had instrument false-positive results than for patients who had instrument true-positive or true-negative results (P = 0.001). We conclude that instrument false positives signaled by the BACTEC 9240 system are not due to bacteria in the blood culture samples.
منابع مشابه
Evaluation of PCR amplification from fixed caecal tissue to determine the infection of experimentally infected laying hens with Brachyspira pilosicoli
Avian intestinal spirochaetosis (AIS) is a condition arising from colonization of the caeca and colon of birds with anaerobic spirochaetes of the genus Brchyspira. The purpose of the present study was to evaluate PCR amplification from formalin-fixed, paraffin-embedded tissue method for detecting infected laying hens with Brachyspira pilosicoli with the same primers that have been used for det...
متن کاملRapid detection of Klebsiella pneumoniae from blood culture bottles by real-time PCR.
A LightCycler real-time PCR hybridization probe-based assay which detects a partial Klebsiella pneumoniae 16S rRNA gene was developed for the rapid identification of K. pneumoniae directly from growth-positive blood culture bottles (BACTEC 9240 system) within 2 h. No cross-reactivity was observed with 65 negative-control blood cultures that grew bacteria other than K. pneumoniae and 48 negative...
متن کاملThe Isolation and Identification of Dominant Lactic Acid Bacteria by the Sequencing of the 16S rRNA in Traditional Cheese (Khiki) in Semnan, Iran
Background: Identification of the dominant lactic acid bacteria involved in the production of traditional cheese in Semnan could be the initiative to protect national genetic resources and produce industrial cheese with desirable texture and organoleptic characteristics similar to traditional cheeses. The present study aimed to determine the biochemical, physiological, and phenotypic properties...
متن کاملMulticenter comparison of BACTEC 9050 and BACTEC 9240 blood culture systems.
The overall recovery of organisms and time to detection with the BACTEC 9050 and BACTEC 9240 systems were compared in a multicenter evaluation. In the first phase of the study, a total of 4,383 compliant aerobic (Plus Aerobic/F) blood culture sets were processed. There was no significant difference in the recovery of individual groups of organisms with the two systems, with the exception of Str...
متن کاملIsolation and identification of Mycoplasma agalactiae by culture and polymerase chain reaction (PCR) from affected sheep to Contagious agalactia of Khuzestan province, Iran
Mycoplasma agalactiae (M. agalactiae) is one of the main causes of contagious agalactia, an infectious syndrome of sheep and goats in Khuzestan province –southwest of Iran that is characterized by mastitis and subsequent failure of milk production, arthritis, abortion and keratoconjunctivitis. This study was carried out to isolation and identification of M. agalactiae with culture and polymeras...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of clinical microbiology
دوره 39 10 شماره
صفحات -
تاریخ انتشار 2001