An evaluation of organ dose modulation on a GE optima CT660‐computed tomography scanner

نویسندگان

  • Matthew T. Dixon
  • Robert J. Loader
  • Gregory C. Stevens
  • Nick P. Rowles
چکیده

Organ Dose Modulation or ODM (GE Healthcare, Milwaukee, WI) was evaluated to characterize changes in CTDIvol, image noise, effective dose, and organ dose saving to patients. Three separate investigations were completed: a tube current modulation phantom was scanned with and without ODM, a CTDIvol phantom was scanned with ODM, and Monte Carlo simulations were performed. ODM was found to reduce the CTDIvol by approximately 20% whilst increasing the noise by approximately 14%. This was reflected in the dose distribution, where the anterior peripheral dose was reduced by approximately 40% whilst the identical poste-rior dose remained largely unaffected. Enabling ODM for the entire scan would reduce the effective dose by approximately 24%; however, this saving reduces to 5% if the images are matched for CTDIvol. These savings mostly originated from reductions in dose to the stomach, breasts, colon, bladder, and liver. ODM has the effect of a global reduction in CTDIvol with an associated increase in image noise. The benefit of ODM was found to be reduced when the dose-saving contribution from the reduced CTDIvol was removed. Given that there is a higher contribution to effective dose throughout the body from the anterior projections, consideration should be given to applying ODM throughout.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Thorax organ dose estimation in computed tomography based on patient CT data using Monte Carlo simulation

Background:  This study presents patient specific and organ dose estimation in computed tomography (CT) imaging of thorax directly from patient CT image using Monte Carlo simulation.  Patient's CT image is considered as the patient specific phantom and the best representative of patient physical index in order to calculate specific organ dose. Materials and Methods: EGSnrc /BEAMnr...

متن کامل

Technical Note: Phantom study to evaluate the dose and image quality effects of a computed tomography organ-based tube current modulation technique.

PURPOSE This technical note quantifies the dose and image quality performance of a clinically available organ-dose-based tube current modulation (ODM) technique, using experimental and simulation phantom studies. The investigated ODM implementation reduces the tube current for the anterior source positions, without increasing current for posterior positions, although such an approach was also e...

متن کامل

Assessment the effect of CT localizer parameter on radiation dose and image quality of GE CT spiral images, A phantom study

Introduction: Using computed tomography (CT) is increasing significantly in Iran and worldwide. Reduction of CT's radiation dose is a critical concern. The aim of this study was evaluating the effect of localizer kVp on spiral CT images when the tube current modulation (TCM) is activated. Materials and Methods: A special wedding cake phantom with circular in...

متن کامل

Dose Reduction Methods in Chest Computed Tomography

Introduction: Computed tomography is one of the most powerful tools for investigating thorax disease, because it shows progression of lung disease much clearer than chest radiography. Technical advances in CT scanners have led to some abilities for CT scanners such as coverage of larger volume, lower noise and ability for acquiring image in one respiratory phase. However, this will be accompani...

متن کامل

Development of CT Scanner Models for Patient Organ Dose Calculations Using Monte Carlo Methods

There is a serious and growing concern about the CT dose delivered by diagnostic CT examinations or image-guided radiation therapy imaging procedures. To better understand and to accurately quantify radiation dose due to CT imaging, Monte Carlo based CT scanner models are needed. This dissertation describes the development, validation, and application of detailed CT scanner models including a G...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016