In vitro evaluation of MR imaging issues at 3T for aneurysm clips made from MP35N: Findings and information applied to 155 additional aneurysm clips.

نویسندگان

  • F G Shellock
  • S Valencerina
چکیده

BACKGROUND AND PURPOSE Aneurysm clips need to be tested at 3T to characterize MR imaging concerns, including magnetic field interactions, MR imaging-related heating, and artifacts. Therefore, we evaluated these risks for aneurysm clips. MATERIALS AND METHODS Three different MP35N aneurysm clips (Codman Slim-Line Aneurysm Clip, straight, blade length 25-mm; Codman Slim-Line Aneurysm Clip Graft, 5-mm diameter x 5-mm width; Codman Slim-Line Aneurysm Clip, reinforcing 30 degrees angle, 6-mm x 18-mm) that represented the largest mass for 155 additional clips made from MP35N were tested. The clips were evaluated at 3T for magnetic field interactions, heating, and artifacts. We studied MR imaging-related heating, placing the clip in a gelled-saline-filled phantom with MR imaging performed by using a transmit/receive radio-frequency body coil at a whole-body average SAR of 3 W/kg for 15 minutes. Artifacts were characterized by using T1-SE and GRE pulse sequences. RESULTS Each aneurysm clip showed relatively minor magnetic field interactions, which would not cause movement in situ. Heating was not excessive (highest temperature change, <1.8 degrees C). Artifacts may create problems if the area of interest is in the same area or close to the aneurysm clip. CONCLUSIONS The results of this investigation demonstrated that it would be acceptable (ie, "MR conditional" using current terminology) for patients with these aneurysm clips to undergo MR imaging at < or =3T. Notably, on the basis of the sizes of the clips that underwent testing, these findings pertain to 155 additional aneurysm clips made from the same material.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Aneurysm clips: evaluation of magnetic field interactions with an 8.0 T MR system.

This study was conducted to evaluate magnetic field interactions for aneurysm clips exposed to an 8.0 T magnetic resonance (MR) system. Twenty-six different aneurysm clips were tested for magnetic field translational attraction (deflection angle test) and torque (qualitative assessment method) using previously described techniques. Six of the specific aneurysm clips (i.e. type, model, blade len...

متن کامل

MR imaging of patients with intracranial aneurysm clips.

Four patients with intracranial aneurysm clips made from a variety of alloys were studied without incidence by MR imaging at field strengths ranging from 0.35 to 0.6 T. Knowledge of the type of alloy used in the manufacturing of an aneurysm clip is important in determining whether the clip will or will not deflect in a magnetic field. Ferromagnetic clips show deflection and torque in a magnetic...

متن کامل

Aneurysm clips: evaluation of magnetic field interactions and translational attraction by use of "long-bore" and "short-bore" 3.0-T MR imaging systems.

BACKGROUND AND PURPOSE The use of 3.0-T MR systems is increasing worldwide. We evaluated magnetic field interactions and translational attraction for 32 aneurysm clips in association with exposure to "long-bore" and "short-bore" 3.0-T MR imaging systems. METHODS Thirty-two different aneurysm clips were evaluated in this investigation. Each aneurysm clip was qualitatively evaluated for magneti...

متن کامل

Artifact quantification and tractography from 3T MRI after placement of aneurysm clips in subarachnoid hemorrhage patients

BACKGROUND The application of advanced 3T MRI imaging techniques to study recovery after subarachnoid hemorrhage (SAH) is complicated by the presence of image artifacts produced by implanted aneurysm clips. To characterize the effect of these artifacts on image quality, we sought to: 1) quantify extent of image artifact in SAH patients with implanted aneurysm clips across a range of MR sequence...

متن کامل

Aneurysm clips made of titanium: magnetic characteristics and artifacts in MR.

PURPOSE To evaluate the magnetic characteristics, artifact formation, and implant safety of titanium aneurysm clips for use in MR imaging. METHODS Aneurysm clips made of titanium alloy TiAl6V4 were tested in a magnetometer to determine their magnetic susceptibility and in a 1.5-T MR imager using both a geometric phantom and an animal model. A commercially available alpha-Phynox clip served as...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • AJNR. American journal of neuroradiology

دوره 31 4  شماره 

صفحات  -

تاریخ انتشار 2010