Arterial and venous contributions to near-infrared cerebral oximetry.
نویسندگان
چکیده
BACKGROUND Cerebral oximetry is a noninvasive bedside technology using near-infrared light to monitor cerebral oxygen saturation (Sco2) in an uncertain mixture of arteries, capillaries, and veins. The present study used frequency domain near-infrared spectroscopy to determine the ratio of arterial and venous blood monitored by cerebral oximetry during normoxia, hypoxia, and hypocapnia. METHODS Twenty anesthetized children aged < 8 yr with congenital heart disease of varying arterial oxygen saturation (Sao2) were studied during cardiac catheterization. Sco2, Sao2, and jugular bulb oxygen saturation (Sjo2) were measured by frequency domain near-infrared spectroscopy and blood oximetry at normocapnia room air, normocapnia 100% inspired O2, and hypocapnia room air. RESULTS Among subject conditions, Sao2 ranged from 68% to 100%, Sjo2 from 27% to 96%, and Sco2 from 29% to 92%. Sco2 was significantly related to Sao2 (y = 0. 85 x -17, r = 0.47), Sjo2 (y = 0.77 x +13, r = 0.70), and the combination (Sco2 = 0.46 Sao2 + 0.56 Sjo2 - 17, R = 0.71). The arterial and venous contribution to cerebral oximetry was 16 +/- 21% and 84 +/- 21%, respectively (where Sco2 = alpha Sao2 + beta Sjo2 with alpha and beta being arterial and venous contributions). The contribution was similar among conditions but differed significantly among subjects (range, approximately 40:60 to approximately 0:100, arterial:venous). CONCLUSIONS Cerebral oximetry monitors an arterial/venous ratio of 16:84, similar in normoxia, hypoxia, and hypocapnia. Because of biologic variation in cerebral arterial/venous ratios, use of a fixed ratio is not a good method to validate the technology.
منابع مشابه
Non-invasive assessment of cerebral oxygenation: A comparison of retinal and transcranial oximetry
BACKGROUND To investigate the correlation between cerebral (SO2-transcranial), retinal arterial (SaO2-retinal) and venous (SvO2-retinal) oxygen saturation as measured by near-infrared spectroscopy (NIRS) and retinal oximetry respectively. METHODS Paired retinal and cerebral oxygen saturation measurements were performed in healthy volunteers. Arterial and venous retinal oxygen saturation and d...
متن کاملCerebral Oximetry – Any Value ?
Cerebral oximetry monitoring, used with the intention of optimizing perioperative outcomes, is increasingly being used in the setting of both cardiac and non-cardiac surgery. Cerebral oximetry had its early beginnings in the 1980s with the work of Jöbsis and colleagues. By using multi-wavelength light sources in the near infrared range, these early investigations demonstrated the potential util...
متن کاملUtility of near-infrared spectroscopy for assessing cerebral oxygen saturation during beach chair position.
(NIRS) and jugular venous bulb oximetry during shoulder surgery in the beach chair position. Their results showed a significant lower jugular venous bulb oxygen saturation during beach chair position in the propofol–remifentanil group compared with the sevoflurane–nitrous oxide group. This difference between the two groups was not found for NIRSmeasured cerebral oxygen saturation, which led the...
متن کاملWireless Monitoring of Liver Hemodynamics In Vivo
Liver transplants have their highest technical failure rate in the first two weeks following surgery. Currently, there are limited devices for continuous, real-time monitoring of the graft. In this work, a three wavelengths system is presented that combines near-infrared spectroscopy and photoplethysmography with a processing method that can uniquely measure and separate the venous and arterial...
متن کاملReal-time continuous monitoring of cerebral blood flow autoregulation using near-infrared spectroscopy in patients undergoing cardiopulmonary bypass.
BACKGROUND AND PURPOSE Individualizing mean arterial blood pressure targets to a patient's cerebral blood flow autoregulatory range might prevent brain ischemia for patients undergoing cardiopulmonary bypass (CPB). This study compares the accuracy of real-time cerebral blood flow autoregulation monitoring using near-infrared spectroscopy with that of transcranial Doppler. METHODS Sixty adult ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Anesthesiology
دوره 93 4 شماره
صفحات -
تاریخ انتشار 2000