Patent ductus arteriosus closure using Occlutech® Duct Occluder, experience in Port Elizabeth, South Africa

نویسندگان

  • Lungile Pepeta
  • Adele Greyling
  • Mahlubandile Fintan Nxele
  • Zongezile Masonwabe Makrexeni
چکیده

BACKGROUND Percutaneous closure of patent ductus arteriosus (PDA) has become standard therapy. Experience with the Occlutech® Duct Occluder is limited. METHODS Data regarding ductal closure using Occlutech® Duct Occluder were reviewed and prospectively collected. Demographics, hemodynamic and angiographic characteristics, complications, and outcomes were documented. RESULTS From March 2013 to June 2016, 65 patients (43 females and 22 males) underwent percutaneous closure of the PDA using Occlutech® Duct Occluder. The median age of the patients was 11 months (range, 1-454 months) and the median weight was 8.5 kg (range 2.5-78 kg). The mean pulmonary artery median pressure was 27 mmHg (range, 12-100 mmHg) and the QP: Qs ratio median was 1.8 (range, 1-7.5), with a pulmonary vascular resistance mean of 2.7 WU (standard deviation [SD] ±2.1). Thirty-two patients had Krichenko Type A duct (49%); 7, Type C (11%); 4, Type D (6%); and 22, Type E (34%). The ductal size (narrowest diameter at the pulmonic end) mean was 3.5 mm (SD ± 1.9 mm). The screening time mean was 17.3 min (SD ± 11.6). Out of 63 patients with successful closure of the PDA using Occlutech® Duct Occluder, there were 15 patients with small PDAs; 25 with moderate PDAs, and 23 with large PDAs. In one patient, the device dislodged to the descending aorta, and in two patients, to the right pulmonary artery immediately following deployment, with successful percutaneous (two) and surgical (one) retrieval. Complete ductal occlusion was achieved in all 63 patients on day one. CONCLUSION The Occlutech® Duct Occluder is a safe and effective device for closure of ducts in appropriately selected patients.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Ductal closure using the Amplatzer duct occluder type two: experience in Port Elizabeth hospital complex, South Africa

OBJECTIVE To report outcomes in percutaneous ductal closure using the Amplatzer duct occluder type two (ADO II). METHODS Records of patients admitted for percutaneous closure of patent ductus arteriosus (PDA) were reviewed. RESULTS From May 2009 to July 2012, 36 patients were assigned to closure using the ADO II. There were 21 females and 15 males. The median age was 16.5 (2-233) months; me...

متن کامل

Transcatheter closure of the patent ductus arteriosus using an Amplatzer duct occluder in adults.

Patent ductus arteriosus (PDA) is a congenital heart disease that frequently escapes detection until the patient reaches adulthood. Percutaneous closure of the PDA has been established as a safe alternative to surgery, the Amplatzer duct occluder (ADO) is a new device that has been recently reported in paediatric patients. Our initial experience with an of ADO in an adult population appears to ...

متن کامل

Interventional Closure of a Patent Ductus Arteriosus Using an Amplatz Canine Duct Occluder in an Alpaca Cria

A 6-month old female alpaca cria presented to The Ohio State University for evaluation of a cardiac murmur. Echocardiography revealed a left-to-right shunting patent ductus arteriosus, a restrictive left-to-right shunting perimembranous ventricular septal defect, and secondary moderate left atrial and ventricular dilation. Aortic root angiography demonstrated a type IIA patent ductus arteriosus...

متن کامل

[Right patent ductus arteriosus with an ipsilateral aortic arch: percutaneous closure with amplatzer devices].

The association of a right aortic arch with an ipsilateral patent ductus arteriosus is rare, especially when there are no other intracardiac anomalies. We report three female patients aged 26, 35 and 9 years with this combination in whom previous attempts at surgical closure by thoracotomy and sternotomy were unsuccessful and who subsequently underwent successful percutaneous closure of the def...

متن کامل

Transcatheter Closure of Patent Ductus Arteriosus Using the Amplatzer Duct Occluder.

The Amplatzer duct occluder (ADO) (AGA Medical Corp., Golden Valley, MN) is designed to meet criteria for an optimal transcatheter technique for closure of a patent ductus arteriosus. It was developed as an alternative to surgical closure and other interventional techniques. It is not technically complex, allows controlled deployment, is retrievable and repositionable, and has a low risk of emb...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 10  شماره 

صفحات  -

تاریخ انتشار 2017