Electromagnetic Interference from Swimming Pool Generator Current Causing Inappropriate ICD Discharges

نویسندگان

  • Edward Samuel Roberto
  • Thein Tun Aung
  • Atif Hassan
  • Abdul Wase
چکیده

Electromagnetic interference (EMI) includes any electromagnetic field signal that can be detected by device circuitry, with potentially serious consequences: incorrect sensing, pacing, device mode switching, and defibrillation. This is a unique case of extracardiac EMI by alternating current leakage from a submerged motor used to recycle chlorinated water, resulting in false rhythm detection and inappropriate ICD discharge. A 31-year-old female with arrhythmogenic right ventricular cardiomyopathy and Medtronic dual-chamber ICD placement presented after several inappropriate ICD shocks at the public swimming pool. Patient had never received prior shocks and device was appropriate at all regular follow-ups. Intracardiac electrograms revealed unique, high-frequency signals at exactly 120 msec suggestive of EMI from a strong external source of alternating current. Electrical artifact was incorrectly sensed as a ventricular arrhythmia which resulted in discharge. ICD parameters including sensing, pacing thresholds, and impedance were all normal suggesting against device malfunction. With device failure and intracardiac sources excluded, EMI was therefore strongly suspected. Avoidance of EMI source brought complete resolution with no further inappropriate shocks. After exclusion of intracardiac interference, device malfunction, and abnormal settings, extracardiac etiologies such as EMI must be thoughtfully considered and excluded. Elimination of inappropriate shocks is to "first, do no harm."

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Electromagnetic Interference in a Private Swimming Pool

Although current lead design and filtering capabilities have greatly improved, Electromagnetic Interference (EMI) from environmental sources has been increasingly reported in patients with Cardiac Implantable Electronic Device (CIED) [1]. Few cases of inappropriate intracardiac Cardioverter Defibrillator (ICD) associated with swimming pool has been described [2]. Here we present a case of 64 ye...

متن کامل

Electromagnetic Interference with Implantable Cardioverter Defibrillators Causing Inadvertent Shock: Case Report and Review of Current Literature

As the number of patients having implantable cardioverter defibrillator (ICD) devices is increasing, it is important for the physicians and patients to be aware of situations and conditions that can result in interference with normal functioning of these devices. There are multiple cases of malfunction of ICDs reported in literature and it may be of great significance to have an overview of the...

متن کامل

Inappropriate ICD Shocks for Inappropriate Reasons

ICD shocks can result from a variety of etiologies; determining the proper etiology of the inappropriate shock is essential for correction of the problem. Electromagnetic interference (EMI) can mimic cardiac signals and cause inappropriate defibrillator shocks. We present two cases of inappropriate ICD shocks due to EMI and reversal of the proximal and distal DF-1 lead terminals of the ICD lead...

متن کامل

Inappropriate sensing events revealing electrocautery-induced implantable cardioverter–defibrillator lead failure

Electromagnetic interference (EMI) associated with current implantable cardioverter–defibrillator (ICD) generators and leads are less prone to long-term failure. Thus, it has been suggested that routine post-ICD replacement interrogation of the device may not be necessary. We present the case of a patient who underwent ICD replacement during which therapies were inadvertently not turned off, wh...

متن کامل

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report

Implantable cardioverter defibrillators (ICDs) have been shown to reduce mortality in high-risk patients. Despite several advances in programming ICDs, inappropriate shocks persist and continue to be psychologically and physically disturbing. External electromagnetic interference from electrocautery, welding, acupunctures, low-output transcutaneous electric nerve stimulators, and electronic mus...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2017  شماره 

صفحات  -

تاریخ انتشار 2017