Application of AAPM TG 119 to volumetric arc therapy (VMAT)
نویسندگان
چکیده
The purpose of this study was to create AAPM TG 119 benchmark plans for volumetric arc therapy (VMAT) and to compare VMAT plans with IMRT plan data. AAPM TG 119 proposes a set of test clinical cases for testing the accuracy of IMRT planning and delivery system. For these test cases, we generated two treatment plans, the first plan using 7-9 static dMLC IMRT fields and a second plan utilizing one- or two-arc VMAT technique. Dose optimization and calculations performed using 6 MV photons and Eclipse treatment planning system. Dose prescription and planning objectives were set according to the TG 119 goals. Plans were scored based on TG 119 planning objectives. Treatment plans were compared using conformity index (CI) for reference dose and homogeneity index (HI) (for D(5)-D(95)). For test cases prostate, head-and-neck, C-shape and multitarget prescription dose are 75.6 Gy, 50.4 Gy, 50 Gy and 50 Gy, respectively. VMAT dose distributions were comparable to dMLC IMRT plans. Our planning results matched TG 119 planning results. For treatment plans studied, conformity indices ranged from 1.05-1.23 (IMRT) and 1.04-1.23 (VMAT). Homogeneity indices ranged from 4.6%-11.0% (IMRT) and 4.6%-10.5% (VMAT). The ratio of total monitor units necessary for dMLC IMRT to that of VMAT was in the range of 1.1-2.0. AAPM TG 119 test cases are useful to generate VMAT benchmark plans. At preclinical implementation stage, plan comparison of VMAT and IMRT plans of AAPM TG 119 test case allowed us to understand basic capabilities of VMAT technique.
منابع مشابه
Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملDosimetric Validation of Volumetric Modulated Arc Therapy (VMAT) Using AAPM TG-119 Benchmark Plans in an Upgraded CLINAC 2100CD for Flattening Filter Free (FFF) Photon Beams
Background: Recently we have upgraded our Varian Clinac 2100CD with a 6MV FFF beam, this upgrade being the first of its kind in our country. Even though the dosimetric characteristics of FFF beams have been reported both in experimental and Monte Carlo studies, application in planning and delivery is complex. The aim of this study was to validate the commissioning of upgraded FFF beams dosimetr...
متن کاملDosimetric validation and clinical implementation of two 3D dose verification systems for quality assurance in volumetric‐modulated arc therapy techniques
A pretreatment quality assurance program for volumetric techniques should include redundant calculations and measurement-based verifications. The patient-specific quality assurance process must be based in clinically relevant metrics. The aim of this study was to show the commission, clinical implementation, and comparison of two systems that allow performing a 3D redundant dose calculation. In...
متن کاملA study on the correlation between plan complexity and gamma index analysis in patient specific quality assurance of volumetric modulated arc therapy.
AIM To evaluate the new Octavius 4D system for patient specific quality assurance and to study the correlation between plan complexity and gamma index analysis in patient specific quality assurance of VMAT using the Octavius 4D system. BACKGROUND McNiven (2010) proposed a study to evaluate the utility of a complexity metric, the Modulation Complexity Score, to evaluate the relationship of the...
متن کاملQuantification and modelling of the dosimetric impact of the treatment couch in volumetric modulated arc therapy (VMAT)
Background: As the volumetric modulated arc therapy (VMAT) becoming a main role of treatment ways, the effect of couch top becomes more significant. It is imperative to re-evaluate the couches that previously may have been considered of no importance during early treatment techniques. The impact of couch top on radiation delivery was explored and the couch model was tested with the aim of reduc...
متن کامل