Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone‐beam CT dose estimations
نویسندگان
چکیده
The purpose of this study was to implement full/half bowtie filter models in a com-mercial treatment planning system (TPS) to calculate kilovoltage (kV) cone-beam CT (CBCT) doses of Varian On-Board Imager (OBI) kV X-ray imaging system. The full/half bowtie filter models were created as compensators in Pinnacle TPS using MATLAB software. The physical profiles of both bowtie filters were imported and hard-coded in the MATLAB system. Pinnacle scripts were written to import bowtie filter models into Pinnacle treatment plans. Bowtie filter-free kV X-ray beam models were commissioned and the bowtie filter models were validated by analyzing the lateral and percent-depth-dose (PDD) profiles of anterior/posterior X-ray beams in water phantoms. A CT dose index (CTDI) phantom was employed to calculate CTDI and weighted CTDI values for pelvis and pelvis-spotlight CBCT protocols. A five-year-old pediatric anthropomorphic phantom was utilized to evaluate absorbed and effective doses (ED) for standard and low-dose head CBCT protocols. The CBCT dose calculation results were compared to ion chamber (IC) and Monte Carlo (MC) data for the CTDI phantom and MOSFET and MC results for the pediatric phantom, respectively. The differences of lateral and PDD profiles between TPS calculations and IC measurements were within 6%. The CTDI and weighted CTDI values of the TPS were respectively within 0.25 cGy and 0.08 cGy compared to IC measurements. The absorbed doses ranged from 0 to 7.22 cGy for the standard dose CBCT and 0 to 1.56 cGy for the low-dose CBCT. The ED values were found to be 36-38 mSv and 7-8 mSv for the standard and low-dose CBCT protocols, respectively. This study demonstrated that the established full/half bowtie filter beam models can produce reasonable dose calculation results. Further study is to be performed to evaluate the models in clinical situations.
منابع مشابه
Usability assessment of cone beam computed tomography with a full-fan mode bowtie filter compared to that with a half-fan mode bowtie filter
Background: In intensity modulated radiation therapy, cone beam computed tomography (CT) has been used to evaluate patients prior to treatment. This study conducted a comparative evaluation of the image reconstruction ability of the clinically used half-fan bowtie filter and the full-fan bowtie filter. Materals and Methods: A CT simulation marker was inserted inside a human phantom, and the pel...
متن کاملCommissioning kilovoltage cone‐beam CT beams in a radiation therapy treatment planning system
The feasibility of accounting of the dose from kilovoltage cone-beam CT in treatment planning has been discussed previously for a single cone-beam CT (CBCT) beam from one manufacturer. Modeling the beams and computing the dose from the full set of beams produced by a kilovoltage cone-beam CT system requires extensive beam data collection and verification, and is the purpose of this work. The be...
متن کاملstimate bowtie filter shape in PET/CT scan with TLD
Introduction: The CT machine utilizes a bowtie filter to shape the X-ray beam and remove lower energy photons. Configuration of this bowtie filter is complex and its geometry is often not available in detail. It causes the CT dose index (CTDI) be with the different values in measurement versus Monte Carlo simulation studies and other analytical calculations. It is important esp...
متن کاملEstimate bowtie filter shape in PET/CT scan with TLD
Introduction: The CT machine utilizes a bowtie filter to shape the X-ray beam and remove lower energy photons. Configuration of this bowtie filter is complex and its geometry is often not available in detail. It causes the CT dose index (CTDI) be with the different values in measurement versus Monte Carlo simulation studies and other analytical calculations. It is important esp...
متن کاملInclusion of the dose from kilovoltage cone beam CT in the radiation therapy treatment plans.
PURPOSE Cone beam CT is increasingly being used for daily patient positioning verification during radiation therapy treatments. The daily use of CBCT could lead to accumulated patient doses higher than the older technique of weekly portal imaging. There have been several studies focusing on measurement or calculation of the patient dose from CBCT recently. METHODS This study investigates the ...
متن کامل