A biomechanical evaluation of orthopaedic implants for hip fractures by finite element analysis and in-vitro tests.
نویسندگان
چکیده
The aim of this study was to test the hypothesis that a reinforced gamma nail for the fixation of subtrochanteric fractures would experience less stress during loading compared with a common gamma nail. The issue of whether the use of the stronger implant would result in more stress shielding in the surrounding bone was also addressed. A finite element analysis (FEA) of a synthetic bone was employed to calculate the stress distribution in implant and bone for two fracture types (AO 31-A3.1 and AO 31-A3.3). The FEA was validated by mechanical tests on six synthetic femurs. To test the hypothesis in vitro, mechanical tests on six pairs of fresh-frozen human femurs were conducted. The femurs were supplied with a common or a reinforced gamma nail in a cross-over study design. Strains were measured on the nail shaft to quantify the loading of the nail. The FEA resulted in 3-51 per cent lower stresses for the reinforced gamma nail. No increase in stress shielding could be observed. In the in-vitro tests, the reinforced gamma nail experienced less strain during loading (p < 0.016). The study demonstrated the benefit of a reinforced gamma nail in subtrochanteric fractures. It experienced less stress but did not result in more stress shielding.
منابع مشابه
Effect of Coating Materials on the Fatigue Behavior of Hip Implants: A Three-dimensional Finite Element Analysis
This study aims to validate, using finite element analysis (FEA), the design concept by comparing the fatigue behavior of hip implant stems coated with composite (carbon/PEEK) and polymeric (PEEK) coating materials corresponding to different human activities: standing up, normal walking and climbing stairs under dynamic loadings to find out which of all these models have a better performance in...
متن کاملFinite Element Analysis for the Treatment of Proximal Femoral Fracture
Dynamic hip screw and gamma nail have been widely used to treat the patients with proximal femoral fractures, but clinical failures of those implants are still to be found. This study developed three-dimensional finite element models to investigate the biomechanical performances of the implants. Two kinds of commercially available implants (dynamic hip screw and gamma nail) and one newly design...
متن کاملFinite Element Analysis of Different Hip Implant Designs along with Femur under Static Loading Conditions
Background: The hip joint is the largest joint after the knee, which gives stability to the whole human structure. The hip joint consists of a femoral head which articulates with the acetabulum. Due to age and wear between the joints, these joints need to be replaced with implants which can function just as a natural joint. Since the early 19th century, the hip joint arthroplasty has evolved, a...
متن کاملIntra-operative monitoring of orthopaedic implant stability by vibration analysis
The success rate of orthopaedic implants is strongly dependent on the initial fixation, but the objective intra-operative assessment of primary stability of cementless implants is still a challenge and, moreover, excessive press-fitting of the implant can cause intraoperative fractures. In a series of in vitro and in vivo experiments the change in the frequency response function (FRF) of the hi...
متن کاملOn the Displacement-Stress Continuous Finite Elements
For the analysis of composite media, three different compatible and mixed finite element formulations are presented which apriori enforce the continuity of stresses as well as displacements at the element interfaces. The formulations are applied for the analysis of hi-material interfaces in two problems often encountered in the field of orthopaedic biomechanics, that is the fixation analysis in...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine
دوره 224 10 شماره
صفحات -
تاریخ انتشار 2010