Electrical exposure risk associated with hands-on defibrillation.

نویسندگان

  • Daniel L Lemkin
  • Michael D Witting
  • Michael G Allison
  • Ali Farzad
  • Michael C Bond
  • Mark A Lemkin
چکیده

BACKGROUND The use of hands-on defibrillation (HOD) to reduce interruption of chest compression after cardiac arrest has been suggested as a means of improving resuscitation outcomes. The potential dangers of this strategy in regard to exposing rescuers to electrical energy are still being debated. This study seeks to determine the plausible worst-case energy-transfer scenario that rescuers might encounter while performing routine resuscitative measures. METHODS Six cadavers were acquired and prepared for defibrillation. A custom instrumentation-amplifier circuit was built to measure differential voltages at various points on the bodies. Several skin preparations were used to determine the effects of contact resistance on our voltage measurements. Resistance and exposure voltage data were acquired for a representative number of anatomic landmarks and were used to map rescuers' voltage exposure. A formula for rescuer-received dose (RRD) was derived to represent the proportion of energy the rescuer could receive from a shock delivered to a patient. We used cadaver measurements to estimate a range of RRD. RESULTS Defibrillation resulted in rescuer exposure voltages ranging from 827V to ∼200V, depending on cadaver and anatomic location. The RRD under the test scenarios ranged from 1 to 8J, which is in excess of accepted energy exposure levels. CONCLUSIONS HOD using currently available personal protective equipment and resuscitative procedures poses a risk to rescuers. The process should be considered potentially dangerous until equipment and techniques that will protect rescuers are developed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Hands-On Defibrillation Has the Potential to Improve the Quality of Cardiopulmonary Resuscitation and Is Safe for Rescuers—A Preclinical Study

BACKGROUND Recently, it has been demonstrated that rescuers could safely provide a low, static downward force in direct contact with patients during elective cardioversion. The purpose of our experimental study was to investigate whether shock delivery during uninterrupted chest compressions may have an impact on cardiopulmonary resuscitation (CPR) quality and can be safely performed in a reali...

متن کامل

An Analysis of Electrical Current Flow Through Rescuers in Direct Contact With Patients During Biphasic External Defibrillation

Direct Contact With Patients During Biphasic External Defibrillation Hands-On Defibrillation : An Analysis of Electrical Current Flow Through Rescuers in Print ISSN: 0009-7322. Online ISSN: 1524-4539 Copyright © 2008 American Heart Association, Inc. All rights reserved. is published by the American Heart Association, 7272 Greenville Avenue, Dallas, TX 75231 Circulation doi: 10.1161/CIRCULATIONA...

متن کامل

Hands-on defibrillation and electrocardiogram artefact filtering technology increases chest compression fraction and decreases peri-shock pause duration in a simulation model of cardiac arrest.

BACKGROUND Reducing pauses during cardiopulmonary resuscitation (CPR) compressions result in better outcomes in cardiac arrest. Artefact filtering technology (AFT) gives rescuers the opportunity to visualize the underlying electrocardiogram (ECG) rhythm during chest compressions, and reduces the pauses that occur before and after delivering a shock. We conducted a simulation study to measure th...

متن کامل

Letter by Sullivan regarding article, "Hands-on defibrillation: an analysis of electrical current flow through rescuers in direct contact with patients during biphasic external defibrillation".

BACKGROUND Brief interruptions in chest compressions reduce the efficacy of resuscitation from cardiac arrest. Interruptions of this type are inevitable during hands-off periods for shock delivery to treat ventricular tachyarrhythmias. The safety of a rescuer remaining in contact with a patient being shocked with modern defibrillation equipment has not been investigated. METHODS AND RESULTS T...

متن کامل

Hands-On Defibrillation An Analysis of Electrical Current Flow Through Rescuers in Direct Contact With Patients During Biphasic External Defibrillation

Background—Brief interruptions in chest compressions reduce the efficacy of resuscitation from cardiac arrest. Interruptions of this type are inevitable during hands-off periods for shock delivery to treat ventricular tachyarrhythmias. The safety of a rescuer remaining in contact with a patient being shocked with modern defibrillation equipment has not been investigated. Methods and Results—Thi...

متن کامل