Constancy checks of well‐type ionization chambers with external‐beam radiation units
نویسندگان
چکیده
Constancy checks of a well-type ionization chamber should be performed regularly as part of a quality assurance regime. The goal of this work was to test the feasibility of using a linear accelerator and an orthovoltage unit to check the constancy of a well-type chamber's response to an external radiation source. The reproducibility, linearity with dose, variation with dose-rate, and variation between energy-matched units of the well-type chamber response when exposed to 6 MV beams was examined. The robustness to errors in establishing the measurement conditions, including setting the source-to-surface distance and gantry angle, rotation of the chamber around the central axis of the beam, and the effect of changing the length of the chamber cable exposed to the field, were tested. The reproducibility and linearity with dose of the chamber response, and robustness to errors in establishing the measurement conditions for 100 kVp and 250 kVp beams from an orthovoltage unit, were also examined. The combined uncertainty, including contributions from errors in establishing the reference conditions, for well-type chamber measurements using a 6 MV beam from a linear accelerator is 1.0%. The combined uncertainties for measurements using 100 and 250 kVp beams were 1.8% and 1.5%, respectively. When focus-source distance errors were reduced to ≤ 1 mm, the combined uncertainties for the 100 and 250 kVp beams were 1.2% and 1.1%, respectively, when the dose to the chamber was confined to the linear region of the dose-response curve. The response of a well-type chamber should remain constant to within 1.2% when exposed to a constant dose from an external beam unit, if reference conditions can be reproducibly established. However, the uncertainty for establishing reference conditions for output measurements for an orthovoltage unit can be reduced, which would justify a reduction of the tolerance for constancy measurements.
منابع مشابه
Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
Background: Absolute dosimetry of external beam radiotherapy is carried out by the use of ionization chambers. These chambers must be calibrated at a standard dosimetry laboratory before any use in clinical dosimetry. The secondary standard dosimetry laboratory of Iran (SSDL) has the duty of calibrating the ionization chambers used in radiotherapy centers in Iran. Materials and Methods...
متن کاملUse of a radioactive check device for redundancy check of ionization chambers
The use of a radioactive check device containing Strontium 90 was investigated to carry out the redundancy checks of Farmer-type ionization chambers. It was not possible to meet the recommended tolerance limits of the redundancy checks without taking into account the angular response of the ionization chambers. The ionization chambers exhibited a maximum variation of 1% in the angular response ...
متن کاملSome steps towards establishing a tertiary standard dosimetry laboratory at a radiotherapy department
Background: In order to deliver the precise dose to the target in radiotherapy, absorbed dose to water at the reference point should be assessed. When the calibration procedure is performed for a reference dosimeter in the 60Co beam of a Secondary Standard Dosimetry Laboratory (SSDL), the total uncertainty in absorbed dose to water (Dw) is estimated to be approximately 1.5%. This study attempts...
متن کاملMeasurements of output factors using different ionization chambers and build up caps
Background: The aim of this work was to study the output factors for Linac using different ionization chambers and build up caps. Materials and Methods: Output factors were measured for open square fields (3 × 3 cm to 40 × 40 cm) defined by collimator jaws for 6 and 15 MV photon beams from a Varian Clinac 2100C accelerator were measured. The measurements in air were performed using Com...
متن کاملInterpretation of In-air Output Ratio of Wedged Fields in Different Measurement Conditions
Introduction: The head scatter factor (Sc) is one of the important parameters for monitor unit (MU) calculation. There are multiple factors that impact the Sc values, such as, head structures, back scattering in to dose monitoring chambers, wedges and so on. This study aimed to investigate the variations of SC with different build-up cap materials, wall thickness, Source to ...
متن کامل