Evaluation of a new rebound tonometer for self-measurement of intraocular pressure.
نویسندگان
چکیده
BACKGROUND/AIMS To compare the accuracy of self-obtained, partner-obtained and trainer-obtained measurements using the handheld Icare Home rebound tonometer with Goldmann applanation tonometry (GAT), and to evaluate the acceptability to subjects of Icare Home measurement. METHODS 76 subjects were trained to use Icare Home for self-measurement using a standardised protocol. A prespecified checklist was used to assess the ability of a subject to perform self-tonometry. Accuracy of Icare Home self-measurement was compared with GAT using one eye per subject, randomly selected. Bland-Altman difference analysis was used to compare Icare Home and GAT intraocular pressure (IOP) estimates. Acceptability of self-tonometry was evaluated using a questionnaire. RESULTS 56 subjects (74%, 95% CI 64 to 84) were able to correctly perform self-tonometry. Mean bias (95% limits of agreement) was 0.3 mm Hg (-4.6 to 5.2), 1.1 mm Hg (-3.2 to 5.3) and 1.2 mm Hg (-3.9 to 6.3) for self-assessment, partner-assessment and trainer-assessment, respectively, suggesting underestimation of IOP by Icare Home tonometry. Differences between GAT and Icare Home IOP were greater for central corneal thickness below 500 µm and above 600 µm than data points within this range. Acceptability questionnaire responses showed high agreement that the self-pressure device was easy to use (84%), the reading was quick to obtain (88%) and the measurement was comfortable (95%). CONCLUSIONS Icare Home tonometry can be used for self-measurement by a majority of trained subjects. IOP measurements obtained using Icare Home tonometry by self-assessment and third party-assessment showed slight underestimation compared with GAT.
منابع مشابه
Evaluation of a new rebound tonometer for self-measurement of intraocular pressure Sub-title Self-measurement using Icare Home showed good agreement with Goldmann applanation
Title Evaluation of a new rebound tonometer for self-measurement of intraocular pressure Subtitle Self-measurement using Icare Home showed good agreement with Goldmann applanation tonometry. Icare Home self-measurement was well accepted and reliable results obtained by 3 in 4 subjects, suggesting potential use for home measurement of intraocular pressure. Number of references = 26 Number of sup...
متن کاملEvaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device.
PURPOSE (a) To investigate whether the ICare rebound tonometer can provide accurate measurements of intraocular pressure (IOP) in the hands of an inexperienced user compared with ICare measurements and Goldmann tonometry by a trained technician and (b) to assess the intrauser reproducibility of IOP measurements and the learning curve among patients using the ICare rebound tonometer. METHODS A...
متن کاملComparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.
BACKGROUND To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients. DESIGN A prospective, cross-sectional study was conducted in a private tertiary glaucoma clinic. PARTICIPANTS OR SAMPLES One hundred nine (54 males : 55 females) patients includ...
متن کاملDoes Rebound Tonometry Probe Misalignment Modify Intraocular Pressure Measurements in Human Eyes?
Purpose. To examine the influence of positional misalignments on intraocular pressure (IOP) measurement with a rebound tonometer. Methods. Using the iCare rebound tonometer, IOP readings were taken from the right eye of 36 healthy subjects at the central corneal apex (CC) and compared to IOP measures using the Goldmann applanation tonometer (GAT). Using a bespoke rig, iCare IOP readings were al...
متن کاملA comparison between rebound and Goldmann's tonometers in screening of patients for glaucoma.
Goldmann applanation tonometer (GAT) has been the gold standard for intraocular pressure (IOP) measurement for over five decades. However, the last few years have seen lots of research in an effort to develop the ultimate tonometer. One of the outcomes of that effort is the rebound tonometer (RBT) (Icare; Tiolat Oy, Helsinki, Finland), which has appeared in clinical practice after extensive res...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The British journal of ophthalmology
دوره 100 8 شماره
صفحات -
تاریخ انتشار 2016