MR-compatibility study of a preclinical digital PET/MRI insert
نویسندگان
چکیده
Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany The combination of PET and MRI into a hybrid device is challenging since both systems might interfere with each other. Therefore, the study of the MR-compatibility of such a combined device is crucial to understand and solve potential interference phenomena. Interference problems, which typically occur and are reported by several groups, range from B0 distortion over Signal-To-Noise Ratio (SNR) degradation to gradient disturbances. In this work, we present the current system status and investigations on the interference between a fully digital PET insert and a 3T clinical MRI. The Hyperion-II PET/MR insert is designed to fit into a clinical MRI system and consists of 10 PET modules which are equipped with up to 6 detector stacks whereas one of them employs a 4×4 digital SiPM array for photon detection. For MRI acquisition, the PET insert is equipped with a dedicated PET transparent Tx/Rx mouse RF-coil. To study the influence of the PET insert on the MR performance, SNR measurements using SE sequences as well as dedicated noise scans were performed. The B0 field distortion caused by the PET insert was studied with field mapping and spectroscopic methods. The influence on the PET performance due to MR operation was evaluated by acquiring PET data from different phantoms (point sources and structured phantoms) while applying demanding gradient and RF dominated stress tests as well as normal MR imaging sequences (SE, FFE, EPI and TSE sequences). All experiments were performed with a partly (one of three possible rings) and a fully populated version of the Hyperion-II scanner. The presence of the PET detector causes MR performance degradation on an acceptable level. The same holds true for the PET performance. However, MR stress tests reveal some sensitivity of the PET’s electronic to gradient switching. Details about this study will be presented at the conference.
منابع مشابه
PET/MRI insert using digital SiPMs: Investigation of MR-compatibility
In this work, we present an initial MR-compatibility study performed with the world's first preclinical PET/MR insert based on fully digital silicon photo multipliers (dSiPM). The PET insert allows simultaneous data acquisition of both imaging modalities and thus enables the true potential of hybrid PET/MRI. Since the PET insert has the potential to interfere with all of the MRI's subsystems (s...
متن کاملEvaluation of PET performance and MR compatibility of a preclinical PET/MR insert with digital silicon photomultiplier technology
Department of Physics of Molecular Imaging Systems, RWTH Aachen University, Germany In this work we present detailed characterizations of our preclinical high resolution PET/ MR insert based on the Hyperion-IID platform. The PET/MR insert consists of a ring of 10 singles detection modules, each comprising 2×3 scintillation detector stacks. Each detector stack features a 30×30 pixelated LYSO cry...
متن کاملDevelopment of PET insert for simultaneous PET/MR imaging of human brain
Molecular Imaging Research & Education (MiRe) Laboratory, Department of Electronic Engineering, Sogang University, Seoul, Korea Recently, there has been great interest on the development of combined PET/MR, which is a useful tool for both functional and anatomic imaging. The purpose of this study was to develop a MR compatible PET insert for simultaneous PET and MR imaging of human brain and to...
متن کاملPET performance and MRI compatibility evaluation of a digital, ToF-capable PET/MRI insert equipped with clinical scintillators.
We evaluate the MR compatibility of the Hyperion-II(D) positron emission tomography (PET) insert, which allows simultaneous operation in a clinical magnetic resonance imaging (MRI) scanner. In contrast to previous investigations, this work aims at the evaluation of a clinical crystal configuration. An imaging-capable demonstrator with an axial field-of-view of 32 mm and a crystal-to-crystal spa...
متن کاملPET performance evaluation of a preclinical digital PET/MRI insert
Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany In recent years, several groups have presented combined PET/MRI devices for clinical and preclinical research. Using new technologies like SiPMs, the performance of the systems improved over the years allowing a detailed exploration of the relatively new fiel...