Experimental use of an ultra-low prime neonatal cardiopulmonary bypass circuit utilizing vacuum-assisted venous drainage.
نویسندگان
چکیده
In adult cardiopulmonary bypass surgery, vacuum assisted venous drainage has become a popular technique to augment venous return to the bypass circuit. The application of this technique in neonatal cardiopulmonary bypass surgery could be beneficial to the further miniaturization of neonatal circuitry by coupling radical respositioning of the oxygenator and pump console with decreasing line length. This report communicates the use of an investigational, vacuum assisted venous drainage neonatal circuit that is positioned at patient level utilizing a modified pump console with elevated double head twin roller pumps. The circuit, including the oxygenator, arterial line, venous line, raceway tubing, and a functional level in the venous reservoir has a priming volume of 107 ml. Initial bench and animal tests have demonstrated that this technique may be clinically feasible in CPB applications. With vacuum assisted venous drainage, the goal of asanguinous neonatal cardiac surgery could become a reality. Safety issues must be adequately addressed to ensure that this technique does not impose unacceptable risks.
منابع مشابه
Hematological effects of a low-prime neonatal cardiopulmonary bypass circuit utilizing vacuum-assisted venous return in the porcine model.
Limiting hemodilution in neonates is difficult when extracorporeal circuits require priming volumes that are 2 to 3 times the blood volume of the newborn patient. This extreme hemodilution contributes to the development of significant postbypass coagulation disturbances. The purpose of this project was to design a low-prime neonatal bypass circuit and evaluate the coagulation status after reduc...
متن کاملVacuum-assisted venous drainage and gaseous microemboli in cardiopulmonary bypass.
When conventional gravity siphon venous drainage cannot achieve satisfactory venous drainage during minimally invasive cardiac and neonatal surgeries, assisted venous drainage techniques are needed to ensure adequate flow. One assisted venous drainage technique, vacuum-assisted venous drainage (VAVD), the aid of a vacuum in the venous reservoir, is now widely used to augment venous drainage dur...
متن کاملVacuum-assisted venous drainage, angel or demon: PRO?
Vacuum-assisted venous drainage (VAVD) was proposed to optimize venous drainage during bypass through femoral venous cannulation. It is currently used in both adult and pediatric surgery when siphon gravity venous drainage is suboptimal. In pediatric surgery, the major advantages of VAVD are a significant decrease in cardiopulmonary bypass prime volume and an improved drainage with all collater...
متن کاملAssisted venous drainage, venous air, and gaseous microemboli transmission into the arterial line: an in-vitro study.
The objective of this study was to examine the interaction of cardiopulmonary bypass venous air with assisted venous drainage, focusing on its production of gaseous microemboli in the arterial line. An in-vitro recirculating cardiopulmonary bypass circuit containing fresh whole bovine blood was monitored with a pulsed-doppler microbubble detector. Air of specific amounts was injected into the v...
متن کاملCan vacuum assisted venous drainage be achieved using a roller pump in an emergency? A pilot study using neonatal circuitry.
There has been much advancement in perfusion technology over its 50 years of progression. One of these techniques is vacuum-assisted venous drainage (VAVD). Many perfusionists augment venous drainage using VAVD, typically from a wall vacuum source. This study explores alternates to providing VAVD if the wall vacuum fails. In two porcine laboratories, approximately 36 in. of 3/16-in. tubing was ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The journal of extra-corporeal technology
دوره 30 4 شماره
صفحات -
تاریخ انتشار 1998