Dosimetric impact of tumor treating field (TTField) transducer arrays onto treatment plans for glioblastomas – a planning study
نویسندگان
چکیده
BACKGROUND Tumor-Treating Fields (TTFields) are a novel treatment strategy for glioblastoma (GBM) that is approved for the use concomitantly to adjuvant chemotherapy. Preclinical data suggest a synergistic interaction of TTFields and radiotherapy (RT). However, the dosimetric uncertainties caused by the highly dense arrays have led to caution of applying the TTF setup during RT. METHODS In a RW3 slab phantom we compared the MV- and kV-CT based planned dose with the measured dose. VMAT-plans were optimized on MV-CTs of an Alderson head phantom without TTF arrays and then re-calculated on the same phantom equipped with TTF arrays. Dose at organs at risk (OAR) and target volumes (PTVs) were compared. RESULTS Measurements at a depth of 2, 3 and 4 cm of a RW 3 slab phantom show an attenuation due to TTField arrays of 3.4, 3.7 and 2.7% respectively. This was in-line with calculated attenuations based on MV-CT (1.2, 2.5 and 2.5%) but not with the attenuation expected from kV-CT based calculations (7.1, 8.2 and 8.6%). Consecutive MV-CT based VMAT planning and re-calculation reveals, that the conformity and homogeneity are not affected by the presence of TTField arrays. The dose at organs at risk (OAR) can show increases or decreases by < 0.5 Gy, which should be considered especially in cases next to the scull base. CONCLUSION MV-CT based dose calculation results in reliable dose distributions also in the presence of TTField arrays. There is a small but clinically not relevant interaction between the TTField arrays and VMAT dose application. Thus, daily replacement of TTField arrays is not necessary in regard to deeply located OARs. RT is feasible, when a VMAT treatment plan is optimized to an array free planning CT. As the biologic effect of a concomitant treatment especially on OARs is currently unknown, a concomitant treatment should be performed only within clinical trials.
منابع مشابه
Dosimetric Impact of a Tumor Treating Fields Device for Glioblastoma Patients Undergoing Simultaneous Radiation Therapy
Purpose A recent randomized phase III clinical trial in patients with glioblastoma demonstrated the efficacy of tumor treating fields (TTFields), in which alternating electric fields are applied via transducer arrays to a patient's scalp. This treatment, when added to standard of care therapy, was shown to increase overall survival from 16 to 20.9 months. These results have generated significan...
متن کاملComparison between field-in-field technique and the use of conventional wedges for treatment planning of esophageal cancer
Introduction: This study was conducted to evaluate and quantify the treatment planning performance of MLC-optimized field-in-field planning technique (FIF), also named forward IMRT, versus wedge-based three field (W3F) technique in terms of dosimetric and radiobiological parameters for esophageal carcinoma. Material and Methods: Twenty patients with esophag...
متن کاملComparison and evaluation of dosimetric and radiobiological parameters in two different treatment plans in stomach cancer by use of 3D CRT
Introduction: In 3DCRT the number of therapeutic fields has dramatic impact on treatment results especially in stomach cancer. Traditionally, for radiotherapy of stomach cancer the therapeutic procedure was using two fields (AP- PA or AP Lateral with wedge) to achieve smooth and homogeneous dose distribution in tumor. But nowadays, adding lateral or oblique fields to aforement...
متن کاملAnalysis of Hip Joint Dose in Prostate Cancer Radiation Therapy: A Dosimetric Comparison of Treatment Plans
Background and purpose: Radiation therapy as a part of cancer treatment is used in almost 50-60% of involved cases. In prostate cancer radiation therapy, a large volume of pelvis is irradiated, so, it is necessary to preserve sensitive organs around the treatment area, especially rectum and bladder. In this study, some dosimetric parameters such as minimum dose (Dmin), maximum dose (Dmax), mean...
متن کاملField-In-Field Plan Versus Tangential Wedged Beam Plan in Chest Wall Radiotherapy of Post-Mastectomy Patients: Treatment Planning Study
Introduction: In this study, dose distribution of the chest wall in post-mastectomy breast cancer patients was evaluated and compared in the tangential wedged beam (TWB) and field-in-field (FIF) plans. Materials and Methods: Thirty-six patients with left-sided breast cancer were enrolled in this study. The FIF and TWB plans were generated for each patient to compare dosimetric parameters of th...