Effects of read-out light sources and ambient light on radiochromic film.
نویسندگان
چکیده
Both read-out light sources and ambient light sources can produce a marked effect on coloration of radiochromic film. Fluorescent, helium neon laser, light emitting diode (LED) and incandescent read-out light sources produce an equivalent dose coloration of 660 cGy h(-1), 4.3 cGy h(-1), 1.7 cGy h(-1) and 2.6 cGy h(-1) respectively. Direct sunlight, fluorescent light and incandescent ambient light produce an equivalent dose coloration of 30 cGy h(-1), 18 cGy h(-1) and 0 cGy h(-1) respectively. Continuously on, fluorescent light sources should not be used for film optical density evaluation and minimal exposure to any light source will increase the accuracy of results.
منابع مشابه
Polarization effects on a high-sensitivity radiochromic film.
A new high-sensitivity radiochromic film has been tested for its polarization properties. Gafchromic HS film has been shown to produce a relatively small (less than 3%) variation in the optical density measured at 660 nm wavelength when the light source is fully linear polarized and the film is rotated through a 360 angle. Similar variations are seen when the detector is linearly polarized. If ...
متن کاملFluorescent light effects on FWT-60 radiochromic film.
FWT-60 radiochromic film has been tested for colouration effects from fluorescent light sources and shown to produce a marked colouration when exposed to office fluorescent light sources showing an approximate 1 OD unit per 0.5 J m(-2) exposure to a broad ultraviolet (UV) UVA + UVB spectrum at the peak absorption wavelength. This produces a measurable and quantifiable response to UV exposure. B...
متن کاملUltraviolet radiation dosimetry with radiochromic film.
Radiochromic film is tested for its broad-band response to ultraviolet (UV) B (290-320 nm) and A (320 nm400 nm), visible and infrared radiation produced by a solar simulator and examined for dosimetry in ultraviolet radiation. Results show that MD-55-2 radiochromic film in solar and fluorescent light sources responds almost exclusively to broad-band UVA radiation with negligible colouration fro...
متن کاملDose Distribution Evaluation and Independent Quality Check of Spherical INTRABEAM™ Applicators via Radiochromic EBT2 Film Measurement
Introduction: The present study was conducted to implement a simple practical independent quality check of depth dose and isotropy of the Intrabeam™ therapeutic X-ray machine using radiochromic EBT2 film. Material and Methods: Theindependent quality check of 1.5, 3.5, and 5-cm spherical Intrabeam™ applicators was accomplished using particular EBT2 film cutting pieces with internal rounded edge...
متن کاملSkin Dosimetry in Radiotherapy of Breast Cancer: a Comparison between EBT and EBT3 Radiochromic Films
Objective: Radiochromic EBT3 film is a later generation of radiochromic films. The aim of this study is to compare EBT and EBT3 radiochromic films in radiotherapy fields of breast cancer.Methods: A RANDO phantom was irradiated by a 6 MV Siemens Primus linac with medial and lateral fields of radiotherapy of breast cancer. Dosimetry was performed in various points in the fields using EBT and EBT3...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Physics in medicine and biology
دوره 43 8 شماره
صفحات -
تاریخ انتشار 1998