Comparison of second- and third-generation enzyme immunoassays for detecting antibodies to hepatitis C virus.
نویسندگان
چکیده
Supplemental assays, such as recombinant immunoblot assays (RIBA), are used to confirm detection of antibodies to hepatitis C virus (HCV). However, due to their expense, they are not widely used in developing countries. The purpose of our study was to compare the results of second- and third-generation (G2 and G3, respectively) enzyme immunoassays (EIAs) and to resolve discordant results by using a supplemental assay to assess the reliability of G2 and G3 EIAs to confirm anti-HCV antibody-positive results. We performed both G2 and G3 EIAs for anti-HCV antibodies on 1,134 serum samples collected during the 2nd year of a longitudinal community-based study in Egypt; 35 samples with discordant results were tested by Abbott Laboratories Micro-Particle Immunoassay (M-EIA) and RIBA. Viremia was determined with an in-house nested reverse transcriptase PCR (RT-PCR) to detect HCV RNA. Concordance between the two assays (G2/G3) was 96.9%; 87 (7.7%) samples were positive and 1,012 (89.2%) were negative by both assays. For 17 samples, the discordant results were G2 assay negative and G3 assay positive, and for 18 samples, the discordant results were G2 assay positive and G3 assay negative. Among the 17 G2 assay-negative and G3 assay-positive samples, 15 were M-EIA positive and 7 were PCR positive. Among the 18 G2 assay-positive and G3 assay-negative samples, 2 were M-EIA positive and none were PCR positive. RIBA results from 24 discordant samples showed 87.5% agreement with the G3 EIA, 12.5% agreement with the G2 EIA, and 95.8% agreement with M-EIA. Eleven samples were indeterminate by RIBA and excluded from this analysis. Based on RIBA results, the sensitivity of the G3 EIA was 99%, compared to 89.8% for the G2 EIA, while the specificity of the G3 EIA was 99.8%, compared to 98.9% for the G2 EIA. These results show that the reliability of the G3 EIA in screening these sera is excellent, and the G3 assay can be used in the absence of supplemental tests where resources are limited. RIBA appears not to have advantages over the less expensive M-EIA screening assay. The main disadvantage of RIBA is the occurrence of indeterminate results, especially among problematic samples. Samples giving discordant results in multiple assays are often indeterminate with the RIBA.
منابع مشابه
Evaluation of third-generation assays for detection of anti-hepatitis C virus (HCV) antibodies and comparison with presence of HCV RNA in blood donors reactive to c100-3 antigen.
We tested serum samples from blood donors by using first-, second-, and third-generation enzyme immunoassays or recombinant immunoblot assays. Second- and third-generation assays gave comparable results. Circulating hepatitis C virus RNA was found in a high proportion of reactive samples. A lack of reactivity or low-level reactivity predicted the absence of hepatitis C virus RNA in 100% of the ...
متن کاملPeptide based enzyme immunoassays for detecting hepatitis C antibodies in sera of people at high risk.
AIM To evaluate the performance of three newly introduced enzyme immunoassays (EIAs) for hepatitis C virus (HCV) antibodies, based on synthetic oligopeptides as antigens. METHODS Referred serum samples (n = 173) from people representing groups at high risk of HCV infection were studied. An EIA based on second generation recombinant polypeptide antigens was used for comparison. EIA reactivitie...
متن کاملAnalysis of Immumoreactivity of Heterologously Expressed Non-structural Protein 4B (NS4B) from Hepatitis C Virus (HCV) Genotype 1a
Background: Detection of hepatitis C virus specific antibodies is the initial step in chronic HCV diagnosis. HCV NS4B is among the most immunogenic HCV antigens and has been widely used in commercial Enzyme Immunoassays (EIA). Additionally, NS4B, a key protein in the virus replication, can be an alternative target for antiviral therapy. Objectives: Development of a new method for high-level ex...
متن کاملMulticentric evaluation of new commercial enzyme immunoassays for the detection of immunoglobulin M and total antibodies against hepatitis A virus.
A multicentric clinical study was conducted on representative sera from 1,738 European and U.S. subjects for the evaluation of new anti-hepatitis A virus enzyme immunoassays from Bio-Rad Laboratories. Comparison with reference DiaSorin S.p.A. tests confirmed the good performance of Bio-Rad assays (99.85% and 99.47% overall agreement in detecting total antibodies and IgM, respectively).
متن کاملComparison of Enzyme Immunoassays for Detection of Antibodies to Hepatitis D Virus in Serum.
Serology remains critical for diagnosing hepatitis D virus (HDV) infection, which affects 15 to 20 million people worldwide, but the literature on characterizing commercial enzyme immunoassays (EIAs) dates back to 15 years ago. We evaluated 2 commercial EIAs currently available for detecting anti-HDV antibodies. The DiaSorin assay demonstrated 100% sensitivity and specificity. Using a modified ...
متن کاملEvaluation of a third generation anti-HCV assay in predicting viremia in patients with positive HCV antibodies.
The most practical screening test for hepatitis C virus antibodies are second and third- generation enzyme immunoassays. We evaluated the usefulness of the third generation microparticle enzyme immunoassay (MEIA) in predicting HCV viraemia in anti-HCV positive patients. Serum samples from 106 patients with positive anti-HCV were obtained. To evaluate the diagnostic value of the MEIA test in pre...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of clinical microbiology
دوره 40 5 شماره
صفحات -
تاریخ انتشار 2002