Computerized Biological Brain Phantom for Evaluation of PET and SPECT Reconstruction
نویسندگان
چکیده
A digital brain phantom was created from available primate autoradiographic (AR) data for use in emission computed tomography studies. The tissue was radio-labelled with a functional analogue of the PET agent [18F]-fluoro-deoxyglucose (FDG). Following sacrifice of the animal, film records from serial 20 m thickness sections were digitized and calibrated to obtain ground truth 2D spatial distributions of relative radionuclide density. A 3D version was constructed by using a video subtraction method to align consecutive slices. In order to assess the effects of accurate modelling of activity, the AR data, containing cortical and basal ganglia structures, was used as a phantom in the context of a partialvolume correction method for obtaining accurate regional quantitation. A second phantom, less realistic in terms of activity assignment, was constructed and also tested. The results indicate that quantitation errors due to effects of nonuniform activity in the AR phantom are significant and comparable in magnitude to errors due to non-phantom effects.
منابع مشابه
Influences of reconstruction and attenuation correction in brain SPECT images obtained by the hybrid SPECT/CT device: evaluation with a 3‐dimensional brain phantom
Objective: The aim of this study was to evaluate the influences of reconstruction and attenuation correction on the differences in the radioactivity distributions in 123I brain SPECT obtained by the hybrid SPECT/CT device. Methods: We used the 3-dimensional (3D) brain phantom, which imitates the precise structure of gray mater, white matter and bone regions. It was filled with 123I solution (20...
متن کاملEvaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT Images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators
Objective(s): The purpose of this study was to examine the optimal reconstruction parameters for brain dopamine transporter SPECT images obtained with a fan beam collimator and compare the results with those obtained by using parallel-hole collimators.Methods: Data acquisition was performed using two SPECT/CT devices, namely a Symbia T6 and an Infinia Hawkeye 4 (device A and B) equipped with fa...
متن کاملPerformance Evaluation of FBP Reconstruction in SPECT Imaging
Introduction: The purpose of this study is to define the optimal parameters for the tomographic reconstruction procedure in a routine single photon emission tomography. The Hoffman brain phantom is modified to evaluate the reconstruction method. The phantom was imaged in a 3 and 2-dimensional conformation and the results were compared. Materials and Methods: The 2D phant...
متن کاملA New Approach for Quantitative Evaluation of Reconstruction Algorithms in SPECT
ABTRACT Background: In nuclear medicine, phantoms are mainly used to evaluate the overall performance of the imaging systems and practically there is no phantom exclusively designed for the evaluation of the software performance. In this study the Hoffman brain phantom was used for quantitative evaluation of reconstruction techniques. The phantom is modified to acquire t...
متن کاملEvaluation of iterative reconstruction method and attenuation correction on brain dopamine transporter SPECT using anthropomorphic striatal phantom
Objective(s): The aim of this study was to determine the optimal reconstruction parameters for iterative reconstruction in different devices and collimators for dopamine transporter (DaT) single-photon emission computed tomography (SPECT). The results were compared between filtered back projection (FBP) and different attenuation correction (AC) methods.Methods: An anthropomorphic striatal phant...
متن کامل