A late dissociation of the ceramic component of a sandwich liner cup in a total hip arthroplasty—Case report and review of the literature
نویسندگان
چکیده
We present a case of a 71-year-old woman with unexplained instability of her total hip arthroplasty 12 years after implantation due to loosening of the ceramic component of a sandwich ceramic in polyethylene liner. She experienced in total 2 dislocations in 6 months. These events occurred without fracture of the ceramic liner. Sandwich ceramic liners were introduced with the idea of combining the benefits of ceramic and polyethylene bearings in total hip arthroplasty in order to reduce wear and to allow a ceramic on ceramic bearing to be used in a cup designed for polyethylene inserts only. Although some reports showed promising early results, case reports and later on retrospective analysis showed high fracture rates of the ceramic liner component. We believe that loosening of the ceramic component may precede fracture of the liner and that a ceramic on ceramic configuration is preferable.
منابع مشابه
Fatal Cobalt Cardiomyopathy Following Revision Total Hip Arthroplasty – A Brief Report with Review of Literature
Ceramic bearing surfaces are being increasingly used in young patients undergoing total hip arthroplasty. However,failures have been reported including fractures even with the newer third generation ceramics. The recommendedtreatment for fracture of ceramic bearing surfaces is complete synovectomy and revision total hip arthroplasty.However, disappointing results have also bee...
متن کاملHip Arthroplasty and its Revision in a Child: Case Report and Literature Review
Juvenile idiopathic arthritis is the leading cause of hip replacement in young children. However, arthroplasty in this population is challenging with several concerns about quality of the growing bone, young age for revision surgery, and difficulties in potential several revisions. In this study we introduce a case of a 12-year old who is one of the youngest patients to undergo revision hip art...
متن کاملDissociation of the S-ROM metal-backed polyethylene acetabular liner--a case report and literature review.
The authors report a case of early failure of a total hip arthroplasty with dissociation of the polyethylene liner of an S-ROM Oblong Cup*, which presented as a dislocation. Initial closed reduction of the dislocation was not possible. Open exploration of the hip revealed that the modular Poly Dial* polyethylene acetabular liner had become dislodged from the acetabular cup. The liner was revise...
متن کاملDissociation of Polyethylene Liner of the Dual Mobility Acetabular Component after Closed Reduction of Dislocation: A Case Report
A dual mobility acetabular component has a structure that combines a polyethylene liner and a femoral head, unlike the general design of acetabular cups, making the dissociation of a polyethylene liner highly unlikely. In addition, it increases the range of motion and reduces the possibility of dislocations by increasing a jump distance. A fifty-one-year-old male who had received total hip arth...
متن کاملFracture of Ceramic Liner and Head in a Total Hip Arthroplasty with a Sandwich Type Cup
Due to its advantages, ceramic-on-ceramic bearings have been widely used in young patients for almost 30 years. Long-term survivorship, low wear, and low biological reactivity to particles are some of its characteristics. Even though this material has had a lot of improvements, the risk of fracture is one of the concerns. There have been reports of fracture of ceramic in the acetabular liner an...
متن کامل