Determination of 1,7,7-Trimethyl-bicyclo(2,2,1)heptan-2-one in a Cream Pharmaceutical Formulation by Reversed-phase Liquid Chromatography
نویسندگان
چکیده
A reversed-phase liquid chromatographic method for the determination of 1,7,7-trimethyl-bicyclo(2,2,1)heptan-2-one in a cream formulation is developed and validated. The separation was achieved using an isocratic mobile phase, on a Lichrosorb C8 column. The calibration curve is linear (r(2)= 0.9999) from 25-175% of the analytical concentration of 1.0 mg/ml. The mean percent standard deviation values for intra-day and inter-day precision studies were <1%. The recovery ranges 99.80-100.06% from a cream formulation. The method can be used reliably in quality control for the analysis of bulk cream samples and final product release.
منابع مشابه
A simple, sensitive and rapid isocratic reversed-phase high-performance liquid chromatography method for determination and stability study of curcumin in pharmaceutical samples
Objective: This study was designed to develop and validate a new reversed-phase high-performance liquid chromatography (RP-HPLC) method based on Q2 (R1) International Conference on Harmonization (ICH) guideline for determination of curcumin in pharmaceutical samples. Materials and Methods: The HPLC instrument method was optimized with isocratic elution with acetonitrile: ammonium acetate (45:55...
متن کاملSeparation of Somatropin and Its Degradation Products by High-Performance Liquid Chromatography Using a Reversed-Phase Polymeric Column
The accurate prediction of protein stability is one of the most challenging goals in protein formulation and delivery. In this study, a gradient RP-HPLC method is described for the separation of human growth hormone (hGH) variants as deamidated and oxidized forms. The methodology employed a polymeric poly (styrene-co-divinylbenzene) column and a 1mL/min flow rate of a linear gradient of 0.1% v/...
متن کاملDevelopment and Validation of a Reversed-phase HPLC Method for the Determination of Hydroxybenzene in a Cream Formulation
A rapid, sensitive and specific reversed-phase high performance liquid chromatographic method with diode-array detection has been developed and validated for the determination of hydroxybenzene (0.494%, w/w) in a commercially available cream pharmaceutical formulation. Isocratic chromatography was performed on a C18 column with methanol-water 60:40 (v/v) containing 0.1% phosphoric acid (v/v) as...
متن کاملSeparation of Somatropin and Its Degradation Products by High-Performance Liquid Chromatography Using a Reversed-Phase Polymeric Column
The accurate prediction of protein stability is one of the most challenging goals in protein formulation and delivery. In this study, a gradient RP-HPLC method is described for the separation of human growth hormone (hGH) variants as deamidated and oxidized forms. The methodology employed a polymeric poly (styrene-co-divinylbenzene) column and a 1mL/min flow rate of a linear gradient of 0.1% v/...
متن کاملCultured rat microglial cells synthesize the endocannabinoid 2-arachidonylglycerol, which increases proliferation via a CB2 receptor-dependent mechanism.
Microglia, as phagocytes and antigen-presenting cells in the central nervous system, are activated in such disease processes as stroke and multiple sclerosis. Because peripheral macrophages are capable of producing endocannabinoids, we have examined endocannabinoid production in a macrophage-colony stimulating factor (M-CSF)-dependent rat microglial cell line (RTMGL1) using reversed phase high-...
متن کامل