Evaluation of the dose calculation accuracy in intensity‐modulated radiation therapy for mesothelioma, focusing on low doses to the contralateral lung
نویسندگان
چکیده
This study compares Monte Carlo (MC) with conventional treatment planning system (TPS) calculations. The EGS4nrc MC code, BEAMnrc, was commissioned to simulate a Varian 21Ex Linac. The accuracy of the simulations, including points blocked by the jaws, was evaluated by comparing MC with ion chamber and MOSFET measurements. Eight mesothelioma IMRT cases were planned using Eclipse (pencil beam and superposition convolution algorithms). Dose distributions were recalculated using BEAMnrc/DOSxyz, and compared with TPS. MC agreed with experimental results for IMRT fields within 3% (96% of points). For regions blocked by the jaws, average agreement between MC and experiment was better than 5% up to 20 cm from isocenter. The pencil beam algorithm underestimated lung MLD, V20, and V5, compared with MC, by a mean (range) of 16% (11-22%), 9.0% (2.4-30.1%), and 11.8% (2-30%), respectively. The superposition convolution algorithm gave better agreement of 8.5% (0-17%), 4% (0-12%) and 0% (-6-6%). Mean dose to the targets was better than +/- 5% in all cases. In conclusion, there is excellent correlation between TPS and MC calculations for the target doses. The pencil beam algorithm and superposition convolution algorithms both underestimate lung dose parameters, but the superposition convolution dose offers improvements in dose calculation accuracy for these patients.
منابع مشابه
The evaluation of lung doses for radiation pneumonia risk in stereotactic body radiotherapy: A comparison of intensity modulated radiotherapy, intensity modulated arc therapy, cyberknife and helical tomotherapy
Background: Radiation Pneumonia (RP) is one of the most extensive side effects in Stereotactic Body Radiotherapy (SBRT) of lung cancer. SBRT are performed by means of Intensity Modulated Radiotherapy (IMRT), Intensity Modulated Arc Therapy (IMAT), CyberKnife (CK) or Helical Tomotherapy (HT) treatment methods. In this study, we performed a plan study to determine the plan parameter such as the M...
متن کاملDosimetric comparison of 3-dimensional conformal and intensity-modulated radiotherapy techniques for whole breast irradiation in the prone and supine positions
Background: The aim of this study was to compare the differences of the dosimetric parameters between three-dimensional conformal radiotherapy (3D-CRT) and simultaneous-integrated boost intensity-modulated radiotherapy (SIB-IMRT) techniques in the prone and supine positions for breast irradiation. Materials and Methods: Ten patients underwent a computed tomography simulation in both the prone a...
متن کاملIntensity modulated radiation therapy (IMRT) technique for left breast cancer by different numbers of beam fields
Background: Intensity Modulated Radiotherapy (IMRT) can improve radiotherapy (RT) results by improving healthy tissue sparing. Additionally, IMRT provides more consistent dose deliveries and suppresses secondary tumor formation. RT is a principal treatment in breast cancer (BC). Aim: To evaluate the outcome of the Radiotherapy Plans (RTP) that use IMRT technique to left breast and other organs,...
متن کاملEvaluation of Accuracy and Quality assurance of external beam therapy with photons
Introduction: Receiving exact dose by the patients is vital in radiotherapy. In radiation therapy, the dosimetry of radiations is too important because of successful radiation inquires for delivering the exact dose to the target volume. This study is to evaluate the tolerances and the accuracy of calculated dose of photon beams in the treatment software system. The TECDOC1583 p...
متن کاملInfluence of Compensator Thickness and Field Size on Buildup factor of a Brass Compensator for Intensity- Modulated Radiation Therapy
Introduction: Radiotherapy is one of the effective methods in the treatment of cancer. In the recent methods of radiotherapy, intensity modulated radiotherapy (IMRT) technique by compensators is used. when photon beam passes the compensator, scattered photons can be generated within the compensator. The presence of scattered photons may complicate the design of compensators as...