BACKGROUND Whether all-polyethylene tibial (APT) components are beneficial to patients who received distal femur limb-salvage surgery lacks high-quality clinical follow-up and mechanical evidence. This study aimed to investigate the biomechanics of the distal femur reconstructed with APT tumor knee prostheses using finite element (FE) analysis based on our previous, promising clinical outcome. METHODS Three-dimensional FE models that use APT and metal-backed tibial (MBT) prostheses to reconstruct distal femoral bone defects were developed and input into the Abaqus FEA software version 6.10.1. Mesh refinement tests and gait simulation with a single foot both in the upright and 15°-flexion positions with mechanical loading were conducted. Stress distribution analysis was compared between APT and MBT at the two static positions. RESULTS For both prosthesis types, the stress was concentrated on the junction of the stem and shaft, and the maximum stress in the femoral axis base was more than 100 Mpa. The stress on the tibial surface was relatively distributed, which was 1-19 MPa. The stress on the tibial bone-cement layer of the APT prosthesis was approximately 20 times higher than that on the MBT prosthesis in the same region. The stress on the proximal tibial cancellous bone and cortical bone of the APT prosthesis was 3-5 times greater than that of the MBT prosthesis, and it was more distributed. CONCLUSIONS Although the stress of bone-cement around the APT component is relatively high, the stress was better distributed at the polyethylene-cement-bone interface in APT than in MBT prosthesis, which effectively protects the proximal tibia in distal femur tumor knee prosthesis replacement. These results should be considered when selecting the appropriate tibial component for a patient, especially under the foreseeable conditions of osteoporosis.
منابع مشابه
Limb-salvage treatment of en-block resected distal femoral tumors with endoprosthesis of all-polyethylene tibial component: a 9-year follow-up study
OBJECTIVE To evaluate the medium-term outcome of limb-salvage surgery using all-polyethylene tibial endoprosthetic replacement following en-block resection for distal femoral tumors. METHODS Forty-nine patients with distal femoral tumor were treated between June 2006 and June 2012. The follow-up period was 6-110 months (average 53.4 months). The prosthetic survival was analyzed using the Kapl...
متن کاملEpithelialization Over a Scaffold of Antibiotic-Impregnated PMMA Beads: A Salvage Technique for Open Tibial Fractures with Bone and Soft Tissue Loss When all Else Fails
The management of soft tissue defects in tibial fractures is essential for limb preservation. Current techniques are not without complications and may lead to poor functional outcomes. A salvage method is described using three illustrative cases whereby a combination of flaps and antibiotic-impregnated polymethylmethacrylate beads are employed to fill the bony defect, fight the infection, and...
متن کاملEffectiveness of Minimally Invasive Plate Osteosynthesis (MIPO) on Comminuted Tibial or Femoral Fractures
Background: Comminuted fractures happen frequently due to traumas. Fixation without opening the fracture site,known as minimally invasive plate osteosynthesis (MIPO), has recently become prevalent. This study has beendesigned to assess the outcomes of this treatment for tibial and femoral comminuted fractures.Methods: A total of 60 patients with comminuted femoral or tibial fractures were opera...
متن کاملDistal femur resection with endoprosthetic reconstruction: a long-term followup study.
The distal femur is a common site for primary and metastatic bone tumors and therefore, it is a frequent site in which limb-sparing surgery is done. Between 1980 and 1998, the authors treated 110 consecutive patients who had distal femur resection and endoprosthetic reconstruction. There were 61 males and 49 females who ranged in age from 10 to 80 years. Diagnoses included 99 malignant tumors o...
متن کاملSingle Cut Distal Femoral Varus Osteotomy (SCFO): A Preliminary Study
Background: Genu valgum usually originates from a deformity of distal femur that is often corrected by distal femoralvarus osteotomy. The osteotomy includes both components of angulationcorrection and translation because the siteof osteotomy is not commonly at the apex of deformity. Improvement of patellar tracking not only depends on valguscorrection, but also it may be partially due to centra...