Identifying and Managing a Malpositioned Endotracheal Tube Bite Block in an Orotracheally Intubated Patient
نویسندگان
چکیده
The universal bite block is increasingly used in orotracheally intubated patients. Here, we report a case of pilot tube dysfunction caused by a malpositioned universal bite block in an orotracheally intubated patient. We summarize the key points on identifying and managing a malpositioned universal bite block from this case and literature review.A 74-year-old woman was emergently intubated during an episode of hyperkalemia-related cardiac arrest. A universal bite block was used for fixing the endotracheal tube. After her condition stabilized, ventilator weaning was attempted; however, a positive cuff-leak test result was observed.The cuff-leak test revealed a lack of elasticity of the pilot balloon, which was completely deflated after 2 mL of air was removed. Pilot tube dysfunction was highly suspected. The bite block was slightly pulled out, and 8 mL of air was aspirated from the pilot tube. The patient was successfully extubated without stridor and respiratory distress.Our case highlighted that a malpositioned bite block may obstruct the pilot tube, causing unfavorable consequences. While fixing the bite block on an endotracheal tube, it is crucial to ensure that the takeoff point of the pilot tube is located within the C-notch of the bite block.
منابع مشابه
Endotracheal tube extubation force: adhesive tape versus endotracheal tube holder.
BACKGROUND Adhesive tape is commonly used to secure the endotracheal tube (ETT) in anesthesia and intensive-care settings. OBJECTIVE To determine the force required to extubate when the ETT is secured with adhesive tape or commercially available ETT holders. METHODS We orally intubated a simulation manikin with a standard 8.0-mm inner-diameter ETT, inflated the cuff to 20 cm H(2)O, and mea...
متن کاملPilot balloon malfunction caused by endotracheal tube bite blocker.
A patient’s bite is a frequent cause of endotracheal tube (ETT) obstruction, which affects the volume and pressure delivered by the ventilator. Bite-related ETT obstruction is a frequent cause of ventilator alarm activation, and can cause respiratory failure, negative-pressure pulmonary edema, and ETT tube/pilot tube damage. Clinicians formerly used objects (syringes, tongue depressors, gauze) ...
متن کاملStudy of endotracheal tube cuff pressure and time of measurement among intensive care units patients
Introduction: The pressure in the endotracheal tube cuff must be kept within an optimal range, to make sure tracheal capillaries circulation still stable during positive ventilation. This is essential to avoid oral and gastric stuffing aspiration Objective: The objective of this study was to assess the endotracheal tube cuff pressure and time intervals measurements among intensive care units...
متن کاملA Note regarding Problems with Interaction and Varying Block Sizes in a Comparison of Endotracheal Tubes
A randomized clinical experiment to compare two types of endotracheal tubes utilized a block design where each of the six participating anesthesiologists performed tube insertions for an equal number of patients for each type of tube. Five anesthesiologists intubated at least three patients with each tube type, but one anesthesiologist intubated only one patient per tube type. Overall, one type...
متن کاملA Case of Death Caused by Tracheal Tube Aspiration
INTRODUCTION Airway management, especially outside the operating room, needs meticulous observation in order to avoid certain risks, such as; endotracheal tube (ETT) disloca-tion, esophageal intubation, and unwanted extubation. ETT or tracheostomy dislocation is responsible for one-half of death or brain damage cases in the ICU. Despite appropri-ate fixation of an ETT, the previously mentioned ...